Related Experiment Video
Updated: Mar 30, 2026

A Plate-Based Assay for the Measurement of Endogenous Monoamine Release in Acute Brain Slices
Published on: August 11, 2021
A high-performance liquid chromatography method for the serotonin release assay is equivalent to the radioactive
N K Sono-Koree1, R A Crist1, E L Frank1,2
1ARUP Laboratories Institute for Clinical and Experimental Pathology, Salt Lake City, UT, USA.
Introduction:
The serotonin release assay (SRA) is considered the gold standard laboratory test for heparin-induced thrombocytopenia (HIT). The historic SRA method uses platelets loaded with radiolabeled serotonin to evaluate platelet activation by HIT immune complexes. However, a nonradioactive method is desirable. We report the performance characteristics of a high-performance liquid chromatography (HPLC) SRA method.
Methods:
We validated the performance characteristics of an HPLC-SRA method, including correlation with a reference laboratory using the radioactive method. Serotonin released from reagent platelets was quantified by HPLC using fluorescent detection. Results were expressed as % release and classified as positive, negative, or indeterminate based on previously published cutoffs.
Results:
Serum samples from 250 subjects with suspected HIT were tested in the HPLC-SRA and with the radioactive method. Concordant classifications were observed in 230 samples (92%). Sera from 41 healthy individuals tested negative. Between-run imprecision studies showed standard deviation of <6 (% release) for positive, weak positive, and negative serum pools. Stability studies demonstrated stability after two freeze-thaw cycles or up to a week of refrigeration.
Conclusion:
The HPLC-SRA has robust performance characteristics, equivalent to the historic radioactive method, but avoids the complexities of working with radioactivity.
More Related Videos
12:21Thermostabilization, Expression, Purification, and Crystallization of the Human Serotonin Transporter Bound to S-citalopram
Published on: November 27, 2016
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018