Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
Automated interpretation of home blood pressure assessment (Hy-Result software) versus physician's assessment: a
Nicolas Postel-Vinay1, Guillaume Bobrie, Alan Ruelland
1aHypertension Unit, Assistance Publique-Hôpitaux de Paris, Hôpital Européen Georges Pompidou, F-75015 Paris, France bINSERM UMRS1138, Centre de Recherche des Cordeliers, E22, Université Paris Descartes, Université Pierre et Marie Curie, Paris, France cUnité de Recherche Clinique, Assistance Publique-Hôpitaux de Paris, Hôpital Européen Georges Pompidou.F-75015 Paris, France dUniversité Paris Descartes, Sorbonne Paris Cité, Faculté de Médecine, F-75006 Paris, France eThot e-santé, Paris fCHU de Quebec - Université Laval. Hôtel-Dieu de Québec. Quebec (QC) Canada G1R 2J6 gPole of Cardiovascular Research, Institut de Recherche Expérimentale et Clinique, Université Catholique de Louvain, Brussels, Belgium hDivision of Cardiology, Cliniques Universitaires Saint-Luc, Université Catholique de Louvain, Brussels, Belgium.
Objective:
Hy-Result is the first software for self-interpretation of home blood pressure measurement results, taking into account both the recommended thresholds for normal values and patient characteristics. We compare the software-generated classification with the physician's evaluation.
Design Method:
The primary assessment criterion was whether algorithm classification of the blood pressure (BP) status concurred with the physician's advice (blinded to the software's results) following a consultation (n=195 patients). Secondary assessment was the reliability of text messages.
Results:
In the 58 untreated patients, the agreement between classification of the BP status generated by the software and the physician's classification was 87.9%. In the 137 treated patients, the agreement was 91.9%. The κ-test applied for all the patients was 0.81 (95% confidence interval: 0.73-0.89). After correction of errors identified in the algorithm during the study, agreement increased to 95.4% [κ=0.9 (95% confidence interval: 0.84-0.97)]. For 100% of the patients with comorbidities (n=46), specific text messages were generated, indicating that a physician might recommend a target BP lower than 135/85 mmHg. Specific text messages were also generated for 100% of the patients for whom global cardiovascular risks markedly exceeded norms.
Conclusion:
Classification by Hy-Result is at least as accurate as that of a specialist in current practice (http://www.hy-result.com).
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: