Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
Accuracy of dental torque wrenches
Abstract:
The aim of this in vitro study was to compare the actual torque of 2 manual wrench systems to their stated (target) torque. New spring- (Nobel Biocare USA, LLC) and friction-style (Zimmer Dental, Inc.) manual dental torque wrenches, as well as spring torque wrenches that had undergone sterilization and clinical use, were tested. A calibrated torque gauge was used to compare actual torque to target torque values of 15 and 35 N/cm. Data were statistically analyzed via mixed-effects regression model with Bonferroni correction. At a target torque of 15 N/cm, the mean torque of new spring wrenches (13.97 N/cm; SE, 0.07 N/cm) was significantly different from that of used spring wrenches (14.94 N/cm; SE, 0.06 N/cm; P < 0.0001). However, the mean torques of new spring and new friction wrenches (14.10 N/cm; SE, 0.07 N/cm; P = 0.21) were not significantly different. For torque measurements calibrated at 35 N/cm, the mean torque of new spring wrenches (35.29 N/cm; SE, 0.10 N/cm) was significantly different (P < 0.0001) from the means of new friction wrenches (36.20 N/cm; SE, 0.08 N/cm) and used spring wrenches (36.45 N/cm; SE, 0.08 N/cm). Discrepancies in torque could impact the clinical success of screw-retained dental implants. It is recommended that torque wrenches be checked regularly to ensure that they are performing to target values.
Related Concept Videos
Net Torque Calculations
Torque
Torque can be considered as the rotational counterpart to force. Since forces change the translational...
Angle of Twist: Problem Solving
Accuracy and Precision
Accuracy and Precision
Screw: Problem Solving
To evaluate the...