From bench to bedside: Can the improvements in left ventricular assist device design mitigate adverse events and
Vincenzo Tarzia1, Gabriele Di Giammarco2, Michele Di Mauro2
1Cardiac Surgery, University of Padova, Padova, Italy.
Objective:
In vitro tests demonstrated that the new cone-bearing configuration of the Jarvik 2000 (Jarvik Heart Inc, New York, NY) left ventricular assist device exhibits better hydraulic efficiency than the previous pin-bearing design. We investigated the long-term outcomes of patients who received the Jarvik 2000 left ventricular assist device, depending on bearing design.
Methods:
A retrospective review of prospectively collected data from 18 centers included in the Italian Registry was performed. From May 2008 to September 2013, 99 patients with end-stage heart failure were enrolled. Patients were divided into 2 groups according to their Jarvik 2000 suspending mechanism: Group pin included patients with pin bearings (May 2008 to June 2010), and group cone included patients with newer cone bearings (July 2010 to September 2013). The 2 groups did not differ significantly in terms of baseline characteristics.
Results:
A total of 30 of 39 patients (group pin) and 46 of 60 patients (group cone) were discharged. During follow-up, 6 patients underwent transplantation, and in 1 patient the left ventricular assist device was explanted. The cumulative incidence competing risk of the entire cohort for noncardiovascular-related death was 28% (20%-40%); the cumulative incidence competing risk for cardiovascular-related death was 56% (42%-73%): 71% in group pin versus 26% in group cone (P = .034). The multivariate analyses confirmed that the pin-bearing design was a risk factor for cardiovascular death, along with Interagency Registry for Mechanically Assisted Circulatory Support class. Right ventricular failures and ischemic and hemorrhagic strokes were significantly higher in group pin.
Conclusions:
Patients with the new pump configuration showed a better freedom from cardiovascular death and lower incidence of fatal stroke and right ventricular failure. Further studies are needed to prove the favorable impact of pump-enhanced fluid dynamics on long-term results.