Standardized Model of Ventricular Fibrillation and Advanced Cardiac Life Support in Swine
Published on: January 30, 2020
Minimum oxygen flow needed for vital support during simulated post-cardiorespiratory arrest resuscitation
E Sanz-Sanjosé1, J J Ariño Irujo2, C E Sánchez Martín1
1Servicio de Anestesiología y Reanimación, Hospital Clínico San Carlos, Madrid, España.
This study found that the Mark IV and Revivator Res-Q resuscitators are most effective for achieving target Fraction of Inspired Oxygen (FiO2) levels during post-cardiac arrest care. Using these devices with 10 liters per minute oxygen flow is recommended for optimal patient oxygenation.
Area of Science:
- Cardiology
- Critical Care Medicine
- Biomedical Engineering
Background:
- European Resuscitation Council (ERC) and American Heart Association (AHA) guidelines recommend titrating Fraction of Inspired Oxygen (FiO2) to achieve an oxygen saturation (O2Sat) of ≥ 94% in post-cardiac arrest care.
- Determining the optimal oxygen delivery parameters is crucial for effective resuscitation and patient outcomes.
Purpose of the Study:
- To evaluate the minimum oxygen flow and ventilation time required to achieve specific FiO2 levels (0.32 and 0.80) using different resuscitator devices in a simulated post-cardiac arrest scenario.
Main Methods:
- An experimental setup simulating post-cardiac arrest care was used with an artificial lung.
- Four resuscitator devices (Mark IV, SPUR II, Revivator Res-Q, O-TWO) were tested at various oxygen flow rates (2, 5, 10, 15 lpm).
- Statistical analysis included Bonferroni and Mann-Whitney U tests to compare device performance.
Main Results:
- All tested resuscitators and oxygen flows achieved an FiO2 of ≥ 0.32 within 35 seconds.
- The Mark IV was the only device to reach an FiO2 of 0.80 at lower flows (2 or 5 lpm) within 75 seconds.
- At 15 lpm, Mark IV and Revivator Res-Q reached FiO2 ≥ 0.80 in 35 seconds, significantly faster than SPUR II (50 seconds).
Conclusions:
- For achieving an FiO2 of 0.32, any resuscitator with 2 lpm oxygen flow is adequate, though O-TWO may be slower initially.
- To reach an FiO2 of 0.80, 10 lpm oxygen flow is sufficient, with Mark IV and Revivator Res-Q being the preferred devices.
- Revivator Res-Q and Mark IV are recommended over O-TWO and SPUR II for optimal FiO2 delivery in post-cardiac arrest care.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Cardiopulmonary Resuscitation I: Adult
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...