Related Experiment Videos
Refractive results of hyperopic hydrogel intracorneal lenses in primate eyes
B E McCarey1, M B McDonald, G van Rij
1Department of Ophthalmology, Emory University, Atlanta, GA 30322.
Abstract:
Hyperopic hydrogel intracorneal lenses were successfully implanted into 27 of 33 primate eyes. All eyes were evaluated preoperatively and postoperatively at monthly intervals for clinical appearance and refractive alteration. In a preliminary surgical series, several factors, such as tight sutures and implant design, resulted in a poor refractive yield. The final surgical series used a microkeratome with a pediatric microkeratome ring for smooth interface cuts, interrupted suturing with sufficient tension to align the wound without compression, a suture through the lens to prevent its dislocation, and intraoperative keratometry to reduce postoperative cylinder. The predicted vs measured refractive alteration for a range of 6 to 20 diopters had a correlation coefficient of .95. Keratometry changes correlated to the refractive changes with a coefficient of .97 but understand the change in refraction created by the surgery.