Evaluation of CareSens POCT Devices for Glucose Testing in the Routine Hospital Setting
Kagan Huysal1, Yasemin U Budak1, Hakan Demirci2
1Faculty, Department of Clinical Laboratory, Sevket Yilmaz Education and Research Hospital , Bursa, Turkey .
Introduction:
Glucose meters are used routinely in hospital wards to manage blood glucose levels in patients requiring frequent monitoring of blood glucose.
Objective:
Our institution has 50 POC instruments utilized by diverse population of all ages and medical conditions. The primary objective of our study was to investigate whether all these CareSens glucose meters (I-sense Inc, Seoul, South Korea) results in hospitalized patients during routine clinical care jointly satisfy the specified quality specifications, as defined by Clinical and Laboratory Standards Institute (CLSI) guideline POCT12-A3.
Materials And Methods:
The records of hospitalized patients who underwent simultaneous measures of glucose levels with both glucose meters and a central laboratory analyser between January and June 2013 were retrospectively analysed. We also performed a prospective evaluation of the accuracy of the CareSens glucose Strip.
Results:
Glucose concentrations measured in 840 patients ranged from 1.66 to 31.72 mmol/L The Bland-Altman difference plot between the auto analyser and all the 50 CareSens glucosemeters revealed a mean bias of -2.2%, with analytical biases for the two methods varying from -31.1% to 26.8%. Eighty four percent of the glucose meter's glucose values were within ± 12.5% for values 5.54 mmol/L of the comparative laboratory glucose values and 93% of the results were within 20% of the reference for glucose >4.2 mmol/L and 65% of the results were within 0.8 mmol/L for glucose <4.2 mmol/L.
Conclusion:
CareSens glucose meter readings in hospital settings, especially in hypoglycaemic patients, should be confirmed by central laboratory analysers whenever possible.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessing Blood pressure in the Leg
Preparation:
Holter Monitor: 24-Hour Monitoring
Pulse Assessment Sites
SBAR II: Application of SBAR
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...