Related Experiment Video
Updated: Mar 30, 2026

Novel Percutaneous Approach for Deployment of 3D Printed Coronary Stenosis Implants in Swine Models of Ischemic Heart Disease
Published on: February 18, 2020
Successful Retrieval of Uncoiled Coronary Guidewire Using Simple Balloon Method
Shivanand Patil1, Rangraj Ramalingam2, Jayashree Kharge3
1Associate Professor, Department of Cardiology, Sri Jayadeva Institute of Cardiovascular Sciences & Research , Bangalore, India.
Abstract:
Breakage of angioplasty device in coronary artery can cause panic in the catheterization laboratory. These broken fragments may serve as a nidus for thrombus formation; hence, removal of these fragments becomes mandatory. Since the incidence of guidewire fracture during angioplasty are rarely reported, evidence-based approaches are not available for the management of such incidental conditions. Here, we report an interesting case of entrapment and unravelling of guidewire. We successfully retrieved unravelled guidewire using a noncompliant balloon inflated in the guiding catheter. Subsequently, the procedure was completed successfully with an implantation of a stent in the culprit lesion. We are of opinion that this novel technique is quite easy and less cumbersome than other described techniques reported earlier.
More Related Videos
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization