Objective and subjective image quality with prospectively gated versus ECG-controlled tube current modulation using
Rami Abazid1, Osama Smettei1, Sawsan Sayed1
1Prince Sultan Cardiac Center, Qassim, PSCCQ, Buraydah.
Introduction:
Radiation exposure is one of the major limitations of computed tomographic coronary angiography (CTA). The purpose of this study was to compare the objective and subjective image quality and radiation dose using prospective ECG gating (PGA) versus ECG-controlled tube current modulation (ECTCM) scanning techniques.
Methods:
A prospective, single-center study was performed at Prince Sultan Cardiac Centre, Qassim, Saudi Arabia. A total of 104 patients with low-to- intermediate probability of coronary artery disease (CAD) underwent CTA with either PGA or ECTCM acquisition. PGA was performed during the study period and compared with the last 50 CTAs previously done using ECTCM. A 4-point scale was used to assess the image quality subjectively. Objective image quality was assessed using image signal, noise, and signal-to-noise ratio (SNR).
Results:
Patient's Baseline characteristics were not different between the two scanning protocols. The 4-point score of subjective image quality showed no significant differences between the PGA and ECTCM scans (2.9 ± 0.7, 2.96 ± 0.7, respectively; p = 0.87). The objective image quality showed significantly higher noise and lower SNR with PGA compared with ECTCM (31 ± 9, 27 ± 9, respectively; p < 0.001 for noise) and (15 ± 5, 17 ± 7, respectively; p < 0.001 for SNR), with no statistical difference in the image signal (434 ± 123, 425 ± 103 HU, respectively, p = 0.7). Radiation exposure was significantly lower with PGA than with ECTCM. The dose-length product (DLP) for PGA was 334 ± 130 mGy, compared with 822 ± 286 mGy for the ECTCM. This corresponds to a 59% reduction in radiation exposure (p < 0.0001).
Conclusions:
Although prospective ECG-triggered axial scanning increased image noise, it maintained subjective image quality and was associated with a 59% reduction in radiation exposure when compared with ECTCM.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...