Related Experiment Video
Updated: Mar 30, 2026

A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Diagnostic Reference Levels of CT Radiation Dose in Whole-Body PET/CT
Ngoneh Jallow1, Paul Christian2, John Sunderland3
1Emory University, Atlanta, Georgia.
Unlabelled:
The role of CT in PET/CT imaging includes acquisition techniques for diagnostic, anatomic localization, and attenuation correction purposes. Diagnostic reference levels of the volumetric CT dose index (CTDIvol) are available for dedicated CT procedures on selected body regions, but similar reference levels for whole-body CT used in PET/CT examinations are limited. This work reports CTDIvol values from sites that conduct whole-body oncologic PET/CT examinations and participated in the scanner validation program of the Society of Nuclear Medicine and Molecular Imaging Clinical Trials Network.
Methods:
From 2010 to 2014, a total of 154 sites submitted CT acquisition parameters used in their clinical (18)F-FDG PET/CT oncology protocols. From these parameters, the CTDIvol was estimated using the ImPACT CTDI dosimetry tables. Histograms of CTDIvol values were created for each year, and descriptive statistics, including mean, median, and 75th percentile, were reported. Repeated-measures ANOVA was performed to determine whether significant differences occurred between reporting years.
Results:
A wide range of technical parameters was reported, most notably in tube current. Between 2010 and 2014, the median CTDIvol ranged from 4.9 to 6.2 mGy and the 75th percentile from 9.7 to 10.2 mGy. There was no significant change in CTDIvol between reporting years (repeated-measures ANOVA, P = 0.985).
Conclusion:
The 75th percentile CTDIvol reported in this work was 9.8 mGy averaged over all reporting years. These data provide a resource for establishing CTDIvol reference values specific to performing CT in PET/CT whole-body examinations. The wide ranges of CT acquisition parameters reported by sites suggest that CTDIvol reference levels may be beneficial for optimization of CT protocols.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT