Super-aperture metrology: overcoming a fundamental limit in imaging smooth highly curved surfaces
1Center of Ultra-Precision Optoelectronic Instrument Engineering, Harbin Institute of Technology, No. 92, West Da-Zhi Street, Harbin, 150001, China.
Abstract:
The imaging of smooth, highly curved or tilted surfaces is widely recognized as one of the most challenging and unsolved problems in optical imaging and metrology today. The reason is that even when such surfaces are imaged using high aperture microscope objectives the steepness of the features causes the light to be reflected in such a way that it is not captured by the lens. This is true even in the limiting case of unity numerical aperture since the illuminating light may also be reflected in the forward direction. In order to overcome this fundamental problem we have developed a method whereby such specimens are covered with a readily removable organic fluorescent film thereby creating an isotropic scattering surface. We show that we are readily able to detect slopes with angles close 90° using a 0.75 NA objective--an 82% improvement over the theoretical aperture limit. Issues of variation in film thickness deposition are shown to be readily accommodated. This approach may be used with other fluorophore materials, organic or inorganic, since there is no need for biocompatibility in this application.
More Related Videos
Related Concept Videos
Atomic Force Microscopy
The AFM Probe
The probe is regarded as the heart of any AFM setup and comprises the...
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Bending of Curved Members - Neutral Surface
Consider the curved member described in the previous lesson. According to Hooke's law, which relates stress to strain within the...
Hydrostatic Pressure Force on a Curved Surface