New investigations into the stability of Mesna using LC-MS/MS and NMR
Dahlia Salman1, Julian Swinden2, Jean-Marie R Peron2
1a Researcher in Oncology Pharmacy Practice-Pharmaceutical Analysis, School of Pharmacy and Chemistry, Faculty of Science , Engineering and Computing, Kingston University-London, Kingston upon Thames , London , UK.
Introduction:
It is important for sarcoma patients to receive the correct dose of Mesna as an adjuvant with ifosfamide to reduce the risk of hemorrhagic cystitis. This paper describes a study conducted to evaluate the physicochemical stability of Mesna for injection formulation over 14 days.
Methods:
Mesna samples (n = 4, 20 mg/ml) were incubated in glass vials at 37 + 0.5ºC. Mesna concentrations were determined by liquid chromatography-mass spectrometry (LC-MS/MS), and nuclear magnetic resonance spectroscopy (NMR) was used to detect degradation products. Evaporative losses and pH were also monitored.
Results:
Our results differed from those published in existing literature. Both LC-MS/MS and NMR indicated that Mesna was unstable. The mean percentage decrease in Mesna concentration was 40% by day 14 of the analysis. The presence of Mesna's dimer Dimesna was detected on day 0 and its concentration increased over time. Dimesna was the only by-product identified.
Conclusion:
Both LC-MS/MS and NMR analyses confirmed the instability of Mesna and its conversion into Dimesna.
More Related Videos
09:36Author Spotlight: Discovering New Alkaloids in Plants with Advanced Mass Spectrometry Techniques
Published on: March 8, 2024
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
Related Concept Videos
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Mass Spectrometry: Amine Fragmentation
In amines, the number of nitrogen atoms affects the mass of the molecular ion, which is described by the nitrogen rule of mass spectrometry. This rule states that a compound containing a single...
Mass Spectrometry of Amines
Tandem Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
High-Resolution Mass Spectrometry (HRMS)