A Novel Rescue Technique for Difficult Intubation and Difficult Ventilation
Published on: January 17, 2011
Reduction of Endotracheal Tube Connector Dead Space Improves Ventilation: A Bench Test on a Model Lung Simulating an
1Section of Neonatal-Perinatal Medicine, Department of Pediatrics, University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma. vadim-ivanov@ouhsc.edu.
Background:
The reduction of instrumental dead space is a recognized approach to preventing ventilation-induced lung injury in premature infants. However, there are no published data regarding the effectiveness of instrumental dead-space reduction in endotracheal tube (ETT) connectors. We tested the impact of the Y-piece/ETT connector pairs with reduced instrumental dead space on CO2 elimination in a model of the premature neonate lung.
Methods:
The standard ETT connector was compared with a low-dead-space ETT connector and with a standard connector equipped with an insert. We compared the setups by measuring the CO2 elimination rate in an artificial lung ventilated via the connectors. The lung was connected to a ventilator via a standard circuit, a 2.5-mm ETT, and one of the connectors under investigation. The ventilator was run in volume-controlled continuous mandatory ventilation mode.
Results:
The low-dead-space ETT connector/Y-piece and insert-equipped standard connector/Y-piece pairs had instrumental dead space reduced by 36 and 67%, respectively. With set tidal volumes (VT) of 2.5, 5, and 10 mL, in comparison with the standard ETT connector, the low-dead-space connector reduced CO2 elimination time by 4.5% (P < .05), 4.4% (P < .01), and 7.1% (not significant), respectively. The insert-equipped standard connector reduced CO2 elimination time by 13.5, 25.1, and 16.1% (all P < .01). The low-dead-space connector increased inspiratory resistance by 17.8% (P < .01), 9.6% (P < .05), and 5.0% (not significant); the insert-equipped standard connector increased inspiratory resistance by 9.1, 8.4, and 5.9% (all not significant). The low-dead-space connector decreased expiratory resistance by 6.8% (P < .01) and 1.8% (not significant) and increased it by 1.4% (not significant); the insert-equipped standard connector decreased expiratory resistance by 1.5 and 1% and increased it by 1% (all not significant). The low-dead-space connector increased work of breathing by 4.7% (P < .01), 3.8% (P < .01), and 2.5% (not significant); the insert-equipped standard connector increased it by 0.8% (not significant), 2.5% (P < .01), and 2.8% (P < .01).
Conclusions:
Both methods of instrumental dead-space reduction led to improvements in artificial lung ventilation. Negative effects on resistance and work of breathing appeared minimal. Further testing in vivo should be performed to confirm the lung model results and, if successful, translated into clinical practice.
More Related Videos
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...