In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
Balancing UKA: overstuffing leads to high medial collateral ligament strains
Thomas J Heyse1, Bilal F El-Zayat2, Ronny De Corte3
1Center for Orthopaedics and Trauma Surgery, University Hospital Marburg, Baldingerstrasse, 35043, Marburg, Germany. heyse@med.uni-marburg.de.
Purpose:
Balancing unicondylar knee arthroplasty (UKA) is challenging. If not performed properly, it may lead to implant loosening or progression of osteoarthritis in the preserved compartment. This study was aimed to document the biomechanical effects of improper balancing. We hypothesised that overstuffing would lead to more valgus, higher strain in the medial collateral ligament (sMCL), and higher lateral contact force.
Methods:
Six fresh-frozen cadaver specimens were mounted in a kinematic rig. Three motion patterns were applied with the native knee and following medial UKA (passive motion, open-chain extension, and squatting), while infrared cameras recorded the trajectories of markers attached to femur and tibia. Three inlay thicknesses were tested (8, 9, 10 mm).
Results:
Overstuffed knees were in more valgus and showed less tibial rotation and higher strains in the sMCL (p < 0.05). Lateral contact forces were higher in some specimens and lower in others. Stiffening of the medial compartment by UKA, even well balanced, already leads to a knee more in valgus with a more stressed sMCL. Overstuffing increases these effects. Knees with a tight sMCL may even see lower lateral contact force. Biomechanics were closest to the native knee with understuffing.
Conclusion:
The first two hypotheses were confirmed, but not the latter. This underlines the importance of optimal balancing. Overstuffing should certainly be avoided. Although kinematics is only slightly affected, contact forces and ligament strains are considerably changed and this might be of more clinical importance. It is advisable to use thinner inlays, if stability is not compromised.
Related Concept Videos
Unsymmetric Loading of Thin-Walled Members
The concept of the shear center is crucial in countering the...
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...
Normal Strain under Axial Loading
Eccentric Axial Loading in a Plane of Symmetry
Residual Stresses in Bending
Plastic Deformations of Members with a Single Plane of Symmetry