Isotopic Scintigraphy Coupled With Computed Tomography for the Investigation of Intrathecal Baclofen Device
Flavie Frémondière1, Franck Lacoeuille2, Avigaelle Sher2
1Department of Physical Medicine and Rehabilitation, University of Angers, Angers, France.
Objective:
To assess the potential use of indium-111 diethylenetriamine pentaacetic acid ((111)In-DTPA) scintigraphy coupled with computed tomography (CT) for the investigation of intrathecal baclofen (ITB) device malfunction.
Design:
Retrospective study of a case series of patients.
Setting:
Neurosurgical and physical and rehabilitation medicine departments.
Participants:
Patients (N=7) with reduced ITB effectiveness in whom prior conventional radiographs were inconclusive.
Intervention:
Nine (111)In-DTPA scintigraphic studies and 8 CT scans. Planar acquisitions were followed by tomoscintigraphy combined with CT.
Main Outcome Measure:
Progression of the radiotracer in the pump, catheters, and in the subarachnoid space.
Results:
In 7 cases, scintigraphy coupled with CT showed leakage behind the pump, lack of activity outside the pump reservoir, abrupt interruption of activity in the catheter, or abnormal distribution of the radiotracer, thus demonstrating that the drug did not reach its target. Surgical revision confirmed these findings in 5 cases. In 1 case, combined imagery ruled out device dysfunction. In the remaining case, only planar acquisitions were performed, showing correct diffusion of the radiotracer.
Conclusions:
The combination of scintigraphy and CT provides simultaneous functional and anatomic imagery of the device. The slow infusion of the radioisotope mimics the diffusion of baclofen, and this could be a useful method to explore intrathecal device malfunction. Further studies are required to compare scintigraphy coupled with CT, to radiopaque injection followed by fluoroscopy or CT.
More Related Videos
09:41Imaging CD19+ B Cells in an Experimental Autoimmune Encephalomyelitis Mouse Model using Positron Emission Tomography
Published on: January 20, 2023
06:59Intrathecal Application of a Fluorescent Dye for the Identification of Cerebrospinal Fluid Leaks in Cochlear Malformation
Published on: February 29, 2020
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...