Pulsatile flow index for qualitative measurements of blood flow with duplex ultrasound. An experimental study
C R Cazenave1, K W Sievers, J V Kaude
1Department of Radiology, University of Florida, College of Medicine, Gainesville.
Abstract:
We determined the accuracy of Doppler blood flow measurements in an experimental investigation using a tissue-simulating phantom, pulsatile flow pumps and heparinized blood. A new index for qualitative assessment of blood flow, the pulsed flow index (PFI) is described. The PFI takes advantage of the area under the flow velocity curve between the true zero line and the diastolic baseline. Under conditions of continuous flow, the PFI ranged from 0.82 to 0.94 (mean value 0.90). The PFI was found to be relatively independent of the transducer/vessel angle (+/- 8%) and the inter/intra-operator variation was small (+/- 7.5%, or +/- 7%, respectively).