[Influence of C-Fe Lines Interference Correction on Laser-Induced Breakdown Spectroscopy Measurement of Unburned
Abstract:
In coal-fired plants, Unburned carbon (UC) in fly ash is the major determinant of combustion efficiency in coal-fired boiler. The balance between unburned carbon and NO(x) emissions stresses the need for rapid and accurate methods for the measurement of unburned carbon. Laser-induced breakdown spectroscopy (LIBS) is employed to measure the unburned carbon content in fly ash. In this case, it is found that the C line interference with Fe line at about 248 nm. The interference leads to C could not be quantified independently from Fe. A correction approach for extracting C integrated intensity from the overlapping peak is proposed. The Fe 248.33 nm, Fe 254.60 nm and Fe 272.36 nm lines are used to correct the Fe 247.98 nm line which interference with C 247.86 nm, respectively. Then, the corrected C integrated intensity is compared with the uncorrected C integrated intensity for constructing calibration curves of unburned carbon, and also for the precision and accuracy of repeat measurements. The analysis results show that the regression coefficients of the calibration curves and the precision and accuracy of repeat measurements are improved by correcting C-Fe interference, especially for the fly ash samples with low level unburned carbon content. However, the choice of the Fe line need to avoid a over-correction for C line. Obviously, Fe 254.60 nm is the best
More Related Videos
08:53Dependence of Laser-induced Breakdown Spectroscopy Results on Pulse Energies and Timing Parameters Using Soil Simulants
Published on: September 23, 2013
10:22Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
Related Concept Videos
Atomic Emission Spectroscopy: Interference
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Emission Spectroscopy: Lab
Flame Photometry: Lab
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...