Analytical goals for quantities used to assess thyrometabolic status
1Department of Biochemical Medicine, Ninewells Hospital and Medical School, Dundee, Scotland, UK.
Annals of Clinical Biochemistry
|January 1, 1989
Summary
Establishing analytical goals for thyroid hormone assays, including total T4 and TSH, is crucial for accurate patient monitoring. These targets ensure reliable laboratory results for diagnosing and managing thyroid conditions.
Area of Science:
- Clinical Chemistry
- Endocrinology
- Laboratory Medicine
Background:
- Intra-individual biological variation is key for setting analytical performance goals.
- Existing data on biological variation informs targets for total T4, free T4, total T3, free T3, and TSH.
- For analytes lacking biological variation data, goals can be derived from reference intervals or interlaboratory quality assessment schemes.
Purpose of the Study:
- To define analytical goals for imprecision and inaccuracy in thyroid hormone assays.
- To establish specific performance targets for key thyroid function tests.
- To guide the improvement of laboratory testing for thyroid disorders.
Main Methods:
- Derived analytical goals for imprecision as coefficients of variation (CV) based on intra-individual variation.
- Established stringent CV goals for total T4 monitoring during thyroxine replacement therapy.
- Utilized reference intervals and interlaboratory quality assessment data for analytes lacking biological variation data.
Main Results:
- Recommended CV goals for total T4, free T4, total T3, free T3, and TSH are ≤2.5%, ≤4.7%, ≤5.2%, ≤3.9%, and ≤8.1%, respectively.
- A more stringent CV goal of ≤1.4% is proposed for total T4 in thyroxine replacement therapy monitoring.
- CV goals for reverse T3 and thyroxine-binding globulin were set at ≤10.7% and ≤7.2%, respectively, based on reference values.
Conclusions:
- Achieving defined analytical goals is essential for reliable thyroid hormone measurements.
- Improved analytical performance enhances between-laboratory comparability and supports universal reference values.
- Assay detection limit reporting should be replaced with information on the working range meeting imprecision goals.