Manual versus Mechanical Chest Compressions on Surfaces of Varying Softness with or without Backboards: A Randomized,
Gabriel Putzer1, Anna Fiala1, Patrick Braun1
1Department of Anesthesiology and Critical Care Medicine, Innsbruck Medical University, Innsbruck, Austria.
Background:
Chest compression quality is decisive for overall outcome after cardiac arrest. Chest compression depth may decrease when cardiopulmonary resuscitation (CPR) is performed on a mattress, and the use of a backboard does not necessarily improve compression depth. Mechanical chest compression devices may overcome this problem.
Objectives:
We sought to investigate the effectiveness of manual chest compressions both with and without a backboard compared to mechanical CPR performed on surfaces of different softness.
Methods:
Twenty-four advanced life support (ALS)-certified rescuers were enrolled. LUCAS2 (Physio-Control, Redmond, WA) delivers 52 ± 2 mm deep chest compressions and active decompressions back to the neutral position (frequency 102 min(-1); duty cycle, 50%). This simulated CPR scenario was performed on a Resusci-Anne manikin (Laerdal, Stavanger, Norway) that was lying on 3 different surfaces: 1) a concrete floor, 2) a firm standard mattress, and 3) a pressure-relieving mattress. Data were recorded by the Laerdal Skill Reporting System.
Results:
Manual chest compression with or without a backboard were performed correctly less often than mechanical chest compressions (floor: 33% [interquartile range {IQR}, 27-48%] vs. 90% [IQR, 86-94%], p < 0.001; standard mattress: 32% [IQR, 20-45%] vs. 27% [IQR, 14-46%] vs. 91% [IQR, 51-94%], p < 0.001; and pressure-relieving mattress 29% [IQR, 17-49%] vs. 30% [IQR, 17-52%] vs. 91% [IQR, 87-95%], p < 0.001). The mean compression depth on both mattresses was deeper with mechanical chest compressions (floor: 53 mm [range, 47-57 mm] vs. 56 mm [range, 54-57 mm], p = 0.003; standard mattress: 50 mm [range, 44-55 mm] vs. 51 mm [range, 47-55 mm] vs. 55 mm [range, 54-58 mm], p < 0.001; and pressure-relieving mattress: 49 mm [range, 44-55 mm] vs. 50 mm [range, 44-53 mm] vs. 55 mm [range, 55-56 mm], p < 0.001). In this ∼6-min scenario, the mean hands-off time was ∼15 to 20 s shorter in the manual CPR scenarios.
Conclusions:
In this experimental study, only ∼30% of manual chest compressions were performed correctly compared to ∼90% of mechanical chest compressions, regardless of the underlying surface. Backboard use did not influence the mean compression depth during manual CPR. Chest compressions were deeper with mechanical CPR. The mean hands-off time was shorter with manual CPR.
Related Concept Videos
Cardiopulmonary Resuscitation I: Adult
Cardiopulmonary Resuscitation II: ACLS Airway Management
Flail Chest-II
Assessment:
1. Clinical Evaluation:
History:
Cardiopulmonary Resuscitation III: AED Use
Chest Physiotherapy
Purpose
CPT is primarily used for patients with excessive bronchial secretions who have difficulty clearing...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...