Impact of oxygen sources on performance of the Ventrain(®) ventilation device in an in vitro set-up
A R Schmidt1, K Ruetzler2,3, T Haas1
1Department of Anaesthesia, University Children's Hospital, Zurich, Switzerland.
Background:
The Ventrain(®) (Dolphys Medical, Eindhoven, The Netherlands) is a disposable handheld ventilation device allowing active inspiration and expiration through a transtracheal catheter. This study investigated Ventrain(®) 's performance when used with different clinical oxygen sources in an in vitro set-up.
Methods:
Three anesthesia oxygen sources (wall-mounted flowmeter, respirator oxygen outlet port, and anesthesia ventilator circuit) were used at gas flow rates of 6, 9, 12, and 15 l/min. First, the sources' driving pressures (DP), the insufflation pressure (IP), and the flow at the catheter tip were measured using a gas flow analyzer. Tidal volumes (VT) delivered by the Ventrain(®) were assessed by the ASL5000 test lung (respiratory rate: 15 min(-1), I:E = 1:1, compliance: 100 ml/cmH2O, resistance: 3.06 cmH2O/l/s).
Results:
VT ranged from 82 to 483 ml for inspiration and 82 to 419 ml for expiration. Measured IP, flow, and VT were less dependent on the set gas flow rate but more on the source's DP. With rising DP the IP, the flow at the catheter tip and consequently VT increased. At an approximate target I:E ratio of 1:1, the ratio of inspiratory to expiratory VT increased with higher DP and gas flow rates.
Conclusion:
The oxygen sources resulted in clinically different IP, flows, and VT delivered by the Ventrain(®) at given gas flow rates. This needs to be considered in a clinical emergency situation. Integrating an adjustable pressure valve into Ventrain(®) to allow regulation of the lowest necessary IP would make its use safer.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Bioreactor Controls-II
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...