Clinical evaluation of the Heart Check system, a new quantitative measurement of fresh capillary BNP
Astrid Monfort1,2, Kelly Da Silva3, Nicolas Vodovar1,4
1Inserm UMR-S 942, Paris, France.
Insights
The Alere™ Heart Check (AHC) test offers a viable point-of-care solution for managing heart failure by measuring B-type natriuretic peptide (BNP). While precision is limited, it accurately assesses patient dyspnea and aids in clinical decisions.
Area of Science:
- Cardiology
- Biomarker Analysis
- Point-of-Care Diagnostics
Background:
- Heart failure (HF) management relies on accurate biomarker assessment.
- Rapid, point-of-care (POC) testing can improve clinical decision-making for HF patients.
- B-type natriuretic peptide (BNP) is a key biomarker for diagnosing and managing heart failure.
Purpose of the Study:
- To evaluate the clinical and analytical performance of the Alere™ Heart Check (AHC) test.
- To assess the utility of a rapid POC immunoassay for BNP measurement in fresh capillary whole blood.
- To compare the AHC test with a standard plasma-based BNP assay.
Main Methods:
- Simultaneous BNP measurements were performed using the AHC capillary whole blood test and a plasma-based Abbott architect system.
- 163 patients with stable chronic heart failure were included in the study.
- Performance metrics including correlation, precision, sensitivity, specificity, and predictive values were analyzed.
Main Results:
- A good correlation was observed between the AHC test and the plasma-based assay, with divergence noted at BNP levels >2000 pg/ml.
- The AHC test demonstrated relatively poor precision.
- Despite precision limitations, the AHC test showed equivalent sensitivity, specificity, and predictive values for assessing patient dyspnea and managing heart failure.
Conclusions:
- The Alere™ Heart Check (AHC) BNP test is a suitable point-of-care tool for heart failure management.
- The test provides valuable clinical information for patient assessment, despite its precision limitations.
- AHC facilitates rapid BNP measurement at the point of care, supporting timely clinical decisions.
Aim:
We have evaluated here the clinical and analytical performance of the Alere™ Heart Check (AHC) test, a rapid point-of-care immunoassay designed for the measurement of BNP from fresh capillary whole blood.
Patients & Methods:
One-hundred-and-sixty-three patients with stable chronic heart failure followed at the cardiac rehabilitation were submitted to simultaneous capillary (AHC) and plasma (Abbott architect system) BNP measurements.
Results:
Both methods showed a good correlation, although the values diverged when BNP was higher than 2000 pg/ml. Despite a relatively poor precision of AHC, however, both methods showed the same performances to assess patients' dyspnea and equivalent sensitivity, specificity, negative and positive predicting values.
Conclusion:
AHC BNP test is a good POC for the management of heart failure despite a relatively poor precision.
More Related Videos
08:21Functional Assessment of the Donor Heart During Ex Situ Perfusion: Insights from Pressure-Volume Loops and Surface Echocardiography
Published on: October 11, 2022
10:03Evaluation of Right Ventricular Function in Experimental Models of Pulmonary Arterial Hypertension
Published on: June 27, 2025
Related Concept Videos
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...