Simulation analysis of optimized brachytherapy for uterine cervical cancer: Can we select the best brachytherapy
Ken Yoshida1, Hideya Yamazaki2, Tadayuki Kotsuma3
1Department of Radiation Oncology, Osaka Medical College, Takatsuki, Osaka, Japan.
Purpose:
To choose the optimal brachytherapeutic modality for uterine cervical cancer, we performed simulation analysis.
Methods And Materials:
For each high-risk clinical target volume (HR CTV), we compared four modalities [classical conventional intracavitary brachytherapy (ConvICBT), Image-guided ICBT (IGICBT), intracavitary/interstitial brachytherapy (ICISBT), and interstitial brachytherapy (ISBT) with perineal approach] using dose-volume histograms using eight sizes of HR CTV (2 × 2 × 2 cm to 7 × 4 × 4 cm) and organs at risk model.
Results:
In ConvICBT, the doses covered 90% of the HR CTV [D90(HR CTV)] decreased from 197% prescribed dose (PD) for the HR CTV size (2 × 2 × 2 cm) to 73% PD for 5 × 4 × 4 cm, whereas the other three modalities could achieve 100% PD for all HR CTV sizes. The minimum doses received by the maximally irradiated 2-cm(3) volumes for organs at risks of IGICBT demonstrated lower values than those of ConvICBT for the HR CTV size of 4 × 3 × 3 cm or smaller. ICISBT demonstrated lower values than those of IGICBT for 4 × 3 × 3 cm or larger. ISBT demonstrated lowest values for 5 × 4 × 4 cm or larger.
Conclusions:
HR CTV size of 4 × 3 × 3 cm seems to be a threshold volume in this simulation analysis, and IGICBT is a better choice for smaller HR CTV than the threshold volume. On larger HR CTV, ICISBT or ISBT is the better choice.