Application of a high-throughput, parallel HPLC system for quantitative chiral analysis of pantoprazole
Hao Wang1, Yantong Sun2, Yan Yang1
1College of Life Sciences, Jilin University, Changchun 130012, PR China.
Background:
Chromatographic separation of enantiomers is considered a task in analytical chemistry particularly for high sample throughput. This paper describes a high-throughput parallel HPLC-MS/MS method for the determination of pantoprazole enantiomers.
Results:
Baseline separation of pantoprazole enantiomers was achieved on a Chiralcel OZ-RH column in a run time of 4.5 min. Assays for enantiomers were linear with satisfactory intra- and inter-day precision and accuracy. The assay was suitable for high-throughput analysis as shown by its successful application to a chiral PK study in beagle dog.
Conclusion:
A high-throughput parallel HPLC-MS/MS assay for pantoprazole has been developed and validated. This method provides nearly twofold increased sample throughput, and was shown to be suitable for application in PK studies.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
11:00Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation