Published on: December 30, 2016
Intensity correction for multichannel hyperpolarized 13C imaging of the heart
William Dominguez-Viqueira1, Benjamin J Geraghty1,2, Justin Y C Lau1,2
1Physical Sciences, Sunnybrook Health Sciences Centre, Toronto, ON, Canada.
Purpose:
Develop and test an analytic correction method to correct the signal intensity variation caused by the inhomogeneous reception profile of an eight-channel phased array for hyperpolarized (13) C imaging.
Theory And Methods:
Fiducial markers visible in anatomical images were attached to the individual coils to provide three dimensional localization of the receive hardware with respect to the image frame of reference. The coil locations and dimensions were used to numerically model the reception profile using the Biot-Savart Law. The accuracy of the coil sensitivity estimation was validated with images derived from a homogenous (13) C phantom. Numerical coil sensitivity estimates were used to perform intensity correction of in vivo hyperpolarized (13) C cardiac images in pigs.
Results:
In comparison to the conventional sum-of-squares reconstruction, improved signal uniformity was observed in the corrected images.
Conclusion:
The analytical intensity correction scheme was shown to improve the uniformity of multichannel image reconstruction in hyperpolarized [1-(13) C]pyruvate and (13) C-bicarbonate cardiac MRI. The method is independent of the pulse sequence used for (13) C data acquisition, simple to implement and does not require additional scan time, making it an attractive technique for multichannel hyperpolarized (13) C MRI.
More Related Videos
08:59Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents
Published on: April 15, 2016
14:56Investigating Cardiac Metabolism in the Isolated Perfused Mouse Heart with Hyperpolarized [1-13C]Pyruvate and 13C/31P NMR Spectroscopy
Published on: April 21, 2023
Related Concept Videos
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Imaging Studies for Cardiovascular System IV: CMRI
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...