QRS prolongation after cardiac resynchronization therapy is a predictor of persistent mechanical dyssynchrony
Oguz Karaca1, Onur Omaygenc2, Beytullah Cakal2
1Cardiology Department, Medipol University Faculty of Medicine, Tem Otoyolu Göztepe Çıkışı No:1, Bağcılar, 34214, İstanbul, Turkey. oguzkaraca@hotmail.com.
Purpose:
Prolonged QRS duration is the main selection criterion for cardiac resynchronization therapy (CRT) which ameliorates left ventricular mechanical dyssynchrony (MD). However, consequences of post-CRT QRS prolongation and residual MD have been poorly evaluated. We aimed to define the predictors of persistent MD and hypothesized that CRT-induced QRS change (ΔQRS) might have an impact on residual MD after CRT.
Methods:
A total of 80 patients receiving CRT were included in the study. ΔQRS was calculated as the difference between the baseline and paced QRS intervals. Residual MD was assessed early after device implantation with a longitudinal dyssynchrony index (Yu index). Significant MD was defined as a Yu index ≥ 33 msec. Two groups were created based on residual MD and compared according to clinical, electrocardiographic and echocardiographic features.
Results:
Patients with persistent MD had longer paced QRS durations (182.5 ± 16.2 vs. 165.4 ± 22.5 msec, p = 0.03) and were less likely to have left ventricular (LV) leads located in the posterolateral vein (53 % vs. 85 %, p = 0.002). The linear correlation between the ΔQRS and the Yu index values was modest (Spearman's rho = -0.341, p = 0.002); additionally, a prolonged QRS was strongly associated with MD after CRT (p = 0.00008). Both LV lead localization and CRT-induced QRS prolongation emerged as the significant predictors of persistent MD. A biventricularly paced QRS more than 10 msec longer than the pre-paced QRS width was predictive of persistent MD after CRT (sensitivity = 80 %, specificity = 62 %).
Conclusions:
ΔQRS was found to be associated with residual MD after CRT. Ten milliseconds of QRS prolongation predicted persistent MD after CRT.
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Cardiomyopathy II: Dilated Cardiomyopathy
Dysrhythmias V: Evaluating Dysrhythmias
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Cardiomyopathy IV: Restrictive Cardiomyopathy
Cardiomyopathy III: Hypertrophic Cardiomyopathy