An Application for Pairing with Wearable Devices to Monitor Personal Health Status
Published on: February 3, 2022
Development of a wireless blood pressure measuring device with smart mobile device
İlhan İlhan1, İbrahim Yıldız1, Mehmet Kayrak2
1Department of Mechatronic Engineering, Fac. Eng. Arch., Necmettin Erbakan University, 42090 Konya, Turkey.
A new wireless blood pressure monitor uses smart mobile devices for control and display. This innovative device achieved over 93% accuracy in clinical tests on patients and healthy individuals.
Area of Science:
- Biomedical Engineering
- Mobile Health (mHealth)
- Medical Devices
Background:
- Smart mobile devices are increasingly prevalent, offering powerful hardware and features.
- The potential for integrating smart mobile devices into biomedical applications is significant.
- Existing blood pressure monitoring methods can be cumbersome and less accessible.
Purpose of the Study:
- To develop a wireless blood pressure measuring device integrated with smart mobile devices.
- To utilize smart mobile devices (Android/iOS) as both indicators and control units for the device.
- To explore the feasibility and accuracy of mobile-integrated vital sign monitoring.
Main Methods:
- Developed a wireless blood pressure cuff communicating via Bluetooth with a smart mobile device.
- Implemented a digital filter within the cuff for signal processing, replacing traditional analog circuits.
- Designed a user interface for Android and iOS operating systems for device control and data display.
- Conducted tests on 18 patients and 20 healthy individuals under physician supervision.
Main Results:
- The developed wireless blood pressure device demonstrated high accuracy.
- Achieved an average accuracy of 93.52% in measurements for sick individuals.
- Achieved an average accuracy of 94.53% in measurements for healthy individuals when compared to a sphygmomanometer.
- The smart mobile device interface functioned effectively for control and indication.
Conclusions:
- The wireless blood pressure measuring device integrated with smart mobile technology is a viable and accurate solution.
- This mHealth approach offers a promising alternative for convenient and accessible blood pressure monitoring.
- The use of digital filtering and mobile integration enhances the performance and usability of the device.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Holter Monitor: 24-Hour Monitoring
Errors occurring during blood pressure monitoring
Several factors...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Measurement of Blood Pressure