Optimal boundary detection method and window settings for coronary atherosclerotic plaque volume analysis in coronary

Ran Heo1,2, Hyung-Bok Park1,3, Byoung Kwon Lee4

  • 1Yonsei-Cedars Sinai Integrative Cardiovascular Imaging Research Center, Yonsei University Health System, Seoul, South Korea.

European Radiology
|December 4, 2015
PubMed

Insights

Quantitative plaque volume analysis using coronary CT angiography (QCT) is a new method. The optimal approach uses a fixed 0.3-mm gap and fixed Hounsfield unit (HU) settings for accurate results.

Area of Science:

  • Cardiovascular imaging
  • Medical diagnostics
  • Radiology

Background:

  • Quantitative plaque volume analysis by coronary CT angiography (QCT) is an emerging technique.
  • Accurate plaque volume measurement is crucial for assessing cardiovascular disease.
  • Standardization of QCT methodology is needed for reliable plaque quantification.

Purpose of the Study:

  • To determine the optimal methodology for quantitative plaque volume analysis using coronary CT angiography (QCT).
  • To compare different boundary detection methods and window width/level settings for QCT plaque volume assessment.
  • To validate QCT findings against intravascular ultrasound (IVUS) measurements.

Main Methods:

  • Fifty-one coronary artery segments were analyzed using QCT.
  • Two boundary detection methods were compared: no gap and a fixed 0.3-mm gap.
  • Three window width and level (WW/WL) settings were evaluated: fixed (740/220 HU), segment-adjusted, and aorta-adjusted.

Main Results:

  • The fixed 0.3-mm gap method showed no significant difference compared to IVUS for plaque volume (139.8 ± 93.9 mm³ vs. 136.1 ± 72.8 mm³).
  • The 'no gap' method significantly underestimated plaque volume compared to IVUS (105.4 ± 82.3 mm³ vs. 136.1 ± 72.8 mm³).
  • The fixed HU setting (740/220) demonstrated the least mean difference in plaque volume compared to IVUS (3.8 ± 39.8 mm³).

Conclusions:

  • Optimal quantitative plaque volume assessment using QCT is achieved with a fixed 0.3-mm gap for boundary detection.
  • The fixed Hounsfield unit (HU) setting of 740/220 provides the most accurate plaque volume measurements.
  • This standardized QCT approach enhances the reliability of plaque quantification in coronary artery disease.
Abstract