Related Experiment Video
Updated: Mar 29, 2026

Additive Manufacturing-Enabled Low-Cost Particle Detector
Published on: March 24, 2023
Detection of airborne viruses using electro-aerodynamic deposition and a field-effect transistor
Kyu-Tae Park1, Dong-Guk Cho2, Ji-Woon Park1
1School of Mechanical Engineering, Yonsei University, Seoul 120-749, Korea.
Abstract:
We report a technique for the detection of aerosolized viruses. Conventional field-effect-transistor (FET)-based techniques use solution-based processes, thus require antibody binding to the detection region of the FET prior to the supply of the analyte. With the method described here, virus-antibody-bound particles are delivered to the FET during detection; therefore, neither a pre-treatment antibody binding step on the FET channel nor washing process for virus-antibody-binding are necessary. Our method is based on the concept that virus-antibody-bound particles are larger than the virus or antibody alone, and thus have larger charge numbers following aerosol charging. When these particles are charged by negative ions and electro-aerodynamically deposited on a substrate, there exists a location on the substrate where neither lone virus nor antibody particles land, and where only virus-antibody-bound particles are deposited. If this location coincides with the channel of the FET, the resulting variation in the current can be used to indicate the existence of a virus. By aerosolizing a mixed solution of the virus and the antibody, only the virus-antibody-bound particles were transported to the swCNT-FET, and the electric current in the swCNT-FET decreased to 30% of that measured with no deposited particles.