Is a New Standard Needed for Diffusion Methods for In Vitro Susceptibility Testing of Fosfomycin against Pseudomonas
María Díez-Aguilar1, Laura Martínez-García2, Rafael Cantón3
1Servicio de Microbiología, Hospital Universitario Ramón y Cajal, Instituto Ramón y Cajal de Investigación Sanitaria (IRYCIS), Madrid, Spain Red Española de Investigación en Patología Infecciosa (REIPI), Instituto de Salud Carlos III, Madrid, Spain.
Abstract:
We analyzed fosfomycin susceptibility results in Pseudomonas aeruginosa clinical isolates obtained by MIC gradient strips and disk diffusion methods using two different inocula, 10(8) and 10(6) CFU/ml, and compared them to the agar dilution reference method. Essential and categorical agreements were 93.6% and 95%, respectively, for the 10(6) CFU/ml alternative inoculum, and they were 67.6% and 78.2%, respectively, for the standard inoculum (10(8) CFU/ml). The use of the 10(6) CFU/ml inoculum improves the agreement values and inhibition zone readings.