Renewing the radiopharmaceutical accuracy check service for Canadian radionuclide calibrators
1National Research Council of Canada, 1200 Montreal Road, Building M-35, Ottawa, ON, Canada K1A0R6.
Abstract:
From 1986 to 2000, a fraction of the Canadian nuclear medicine community participated in a service offered by the National Research Council (NRC) of Canada to check the accuracy of administered doses of radiopharmaceuticals. The NRC renewed this service in recent years with a revalidation and reviving of the Secondary Standard Ionizing Radiation Chamber System (SSIRCS). The NRC conducted mock services for (99m)Tc, in varying geometries (syringes and serum vials), on two NRC commercial radionuclide calibrators, and at a nuclear medicine department in Canada. These tests showed the measured doses to be within 10% of the prepared standard but in some cases were sufficiently different from unity to warrant specific geometry factors to be derived.
More Related Videos
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Accuracy and Precision
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Quality Assurance
Data Validation
Key parameters for method validation include:
![Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F55537.jpg&w=3840&q=50)