A sensitive and efficient method for routine pesticide multiresidue analysis in bee pollen samples using gas and
P Parrilla Vázquez1, A Lozano1, S Uclés1
1Agrifood Campus of International Excellence (ceiA3), European Union Reference Laboratory for Pesticide Residues in Fruit & Vegetables, Department of Chemistry and Physics, University of Almeria, La Cañada de San Urbano, 04120 Almería, Spain.
Abstract:
Several clean-up methods were evaluated for 253 pesticides in pollen samples concentrating on efficient clean-up and the highest number of pesticides satisfying the recovery and precision criteria. These were: (a) modified QuEChERS using dSPE with PSA+C18; (b) freeze-out prior to QuEChERS using dSPE with PSA+C18; (c) freeze-out prior to QuEChERS using dSPE with PSA+C18+Z-Sep; and (d) freeze-out followed by QuEChERS using dSPE with PSA+C18 and SPE with Z-Sep. Determinations were made using LC-MS/MS and GC-MS/MS. The modified QuEChERS protocol applying a freeze-out followed by dSPE with PSA+C18 and SPE clean-up with Z-Sep was selected because it provided the highest number of pesticides with mean recoveries in the 70-120% range, as well as relative standard deviations (RSDs) typically below 20% (12.2% on average) and ensured much better removal of co-extracted matrix compounds of paramount importance in routine analysis. Limits of quantification at levels as low as 5μgkg(-1) were obtained for the majority of the pesticides. The proposed methodology was applied to the analysis of 41 pollen bee samples from different areas in Spain. Pesticides considered potentially toxic to bees (DL50<2μg/bee) were detected in some samples with concentrations up to 72.7μgkg(-1), which could negatively affect honeybee health.
More Related Videos
Related Concept Videos
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Types of Detectors-II
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...