Graziella Bonetti1, Vanessa Cancelli1, Giulio Coccoli1
1Clinical Chemistry Laboratory, A.O. Spedali Civili, Brescia, Italy.
Primary Care Diabetes
|December 15, 2015
Summary
Choosing the right blood collection tube is crucial for accurate glucose measurement, as in vitro glycolysis can affect results. GlucoEXACT tubes with sodium fluoride/citrate buffer prevent glycolysis, ensuring reliable glucose analysis even with delayed plasma separation.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Analytical Biochemistry
Background:
- Accurate blood glucose measurement is vital in clinical diagnostics.
- In vitro glycolysis in collected blood samples can lead to inaccurate glucose readings.
- Evaluating different blood collection tubes is necessary to mitigate pre-analytical errors.
Purpose of the Study:
- To compare the suitability of various blood collection tubes for glucose analysis.
- To assess the impact of delayed plasma separation on glucose stability at room temperature.
- To identify the optimal tube for preserving glucose integrity.
Main Methods:
- Blood samples from 49 healthy volunteers were collected in different tube types.
- Reference plasma glucose was immediately processed and stored at room temperature (RT) for up to 4 hours.
- Glucose levels were determined using a hexokinase method.
Main Results:
- Serum with clot activator, lithium heparin, and sodium fluoride/EDTA tubes showed unacceptable glucose bias after 4 hours at RT.
- GlucoEXACT tubes (sodium fluoride/citrate buffer) demonstrated a bias within the desirable limit (±1.8%) at 1, 2, and 4 hours.
- These results indicate superior glucose stability in GlucoEXACT tubes.
Conclusions:
- GlucoEXACT tubes containing sodium fluoride/citrate buffer effectively prevent glycolysis.
- These tubes are equivalent to reference plasma tubes for glucose analysis.
- They allow for up to a 4-hour delay in plasma separation without compromising glucose accuracy.