Comparing the performance of a new disposable pneumatic tocodynamometer with a standard tocodynamometer
Hari Eswaran1, James D Wilson2, Pam Murphy1
1Department of Obstetrics and Gynecology, University of Arkansas for Medical Sciences, Little Rock, AR, USA.
Introduction:
The goal was to test a newly developed pneumatic tocodynamometer (pTOCO) that is disposable and lightweight, and evaluate its equivalence to the standard strain gauge-based tocodynamometer (TOCO).
Material And Methods:
The equivalence between the devices was determined by both mechanical testing and recording of contractile events on women. The data were recorded simultaneously from a pTOCO prototype and standard TOCO that were in place on women who were undergoing routine contraction monitoring in the Labor and Delivery unit at the University of Arkansas for Medical Sciences. In this prospective equivalence study, the output from 31 recordings on 28 pregnant women that had 171 measureable contractions simultaneously in both types of TOCO were analyzed. The traces were scored for contraction start, peak and end times, and the duration of the event was computed from these times.
Results:
The response curve to loaded weights and applied pressure were similar for both devices, indicating their mechanical equivalence. The paired differences in times and duration between devices were subjected to mixed-models analysis to test the pTOCO for equivalence with standard TOCOs using the two-one-sided tests procedure. The event times and duration analyzed simultaneously from both TOCO types were all found to be significantly equivalent to within ±10 s (all p-values ≤0.0001).
Conclusion:
pTOCO is equivalent to the standard TOCO in the detection of the timing and duration of uterine contractions. pTOCO would provide a lightweight, disposable alternative to commercially available standard TOCOs.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Pipe Flowrate Measurement
The orifice meter is a simple,...
Measurement of Blood Pressure