Comparing the performance of a new disposable pneumatic tocodynamometer with a standard tocodynamometer
Hari Eswaran1, James D Wilson2, Pam Murphy1
1Department of Obstetrics and Gynecology, University of Arkansas for Medical Sciences, Little Rock, AR, USA.
A new disposable pneumatic tocodynamometer (pTOCO) is equivalent to standard strain gauge tocodynamometers (TOCO) for measuring uterine contraction timing and duration. This lightweight pTOCO offers a viable alternative for labor monitoring.
Area of Science:
- Biomedical Engineering
- Obstetrics and Gynecology
Background:
- Standard strain gauge tocodynamometers (TOCO) are widely used for uterine contraction monitoring.
- A novel disposable, lightweight pneumatic tocodynamometer (pTOCO) has been developed.
Purpose of the Study:
- To evaluate the equivalence of the new pTOCO to the standard TOCO.
- To assess the performance of the pTOCO in measuring uterine contraction timing and duration.
Main Methods:
- Mechanical testing of both devices under varying loads and pressures.
- Simultaneous recording of uterine contractions using pTOCO and standard TOCO in pregnant women.
- Analysis of 171 contractions from 28 women, focusing on contraction start, peak, end times, and duration.
Main Results:
- Mechanical testing demonstrated similar response curves for both devices.
- Statistical analysis confirmed significant equivalence between pTOCO and standard TOCO for event timing and duration (within ±10 s, p ≤0.0001).
Conclusions:
- The pTOCO is equivalent to the standard TOCO for accurately detecting uterine contraction timing and duration.
- The pTOCO presents a promising lightweight, disposable alternative to existing tocodynamometers for clinical use.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Pipe Flowrate Measurement
The orifice meter is a simple,...
Measurement of Blood Pressure