Assessing the accuracy of a point-of-care analyzer for hyperlipidaemia in western Kenya

Paul H Park1, Patrick Chege2, Isabel C Hagedorn3

  • 1Duke Global Health Institute, Duke University, Durham, NC, USA.

Insights

The CardioChek PA device is inaccurate for measuring cholesterol levels in Kenya, misclassifying patients and posing risks. Validation of diagnostic tools is crucial before widespread use in new settings.

Area of Science:

  • Clinical diagnostics
  • Point-of-care testing
  • Public health in low- and middle-income countries

Background:

  • Hyperlipidaemia prevalence is increasing in low- and middle-income countries.
  • Point-of-care devices like CardioChek PA aim to improve diagnostics by providing rapid results.
  • Validation of such devices in local settings is essential.

Purpose of the Study:

  • To assess the accuracy of the CardioChek PA device for lipid measurement in Kenya.
  • To compare CardioChek PA results against the gold standard laboratory testing.
  • To evaluate the risk of misclassification due to device inaccuracy.

Main Methods:

  • 246 adult subjects in Webuye, Kenya, provided two blood samples.
  • Lipid levels were measured simultaneously using CardioChek PA and laboratory methods.
  • Accuracy was assessed using US National Institutes of Health (NIH) per cent bias parameters.

Main Results:

  • CardioChek PA showed significant inaccuracy for LDL cholesterol (-25.9% bias), HDL cholesterol (-8.2% bias), and total cholesterol (-15.9% bias).
  • Patients at higher risk were more likely to be misclassified into lower risk categories.
  • The device did not meet NIH accuracy recommendations for key lipid parameters.

Conclusions:

  • The CardioChek PA device is inaccurate and unsuitable for clinical use in Kenya.
  • Findings underscore the necessity of validating diagnostic tools in their intended settings before implementation.
  • Potential utility does not negate the need for rigorous local validation.