Ventilator Settings to Avoid Nuisance Alarms During Mouthpiece Ventilation
Related Experiment Video
Updated: Mar 28, 2026

Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Annalisa Carlucci1, Alessio Mattei2, Veronica Rossi3
1Pulmonary Rehabilitation, Fondazione S Maugeri, IRCCS, Pavia, Italy. annalisa.carlucci@fsm.it.
Background:
A recent study found that activation of disconnection and low-pressure alarms is common during mouthpiece ventilation and may represent a major limitation to its use. The aim of this bench study was: (1) to investigate the technical aspects that can influence the setting of the ventilator during mouthpiece ventilation and (2) to provide a practical setting strategy to avoid the alarm activation.
Methods:
Eight life-support ventilators able to deliver volume controlled ventilation were tested in a bench study using a single-limb non-vented circuit configuration connected to a standard mouthpiece. Disconnection and apnea alarm were turned off or set at the least sensitive setting. The backup frequency was set at the lowest available level. Different tidal volumes (VT) (from 500 to 1,200 mL) were tested with the rectangular and descending flow shape. For each VT, we reported the maximum set inspiratory time (TI) that allowed preventing activation of the low-pressure alarm. The presence of auto-triggering was also surveyed.
Results:
We found that a correct combination of VT and TI avoided the activation of disconnection and low-pressure alarms in all but 3 ventilators. One ventilator did not allow mouthpiece ventilation independently from the settings used. The inability to turn off the apnea alarm in two other ventilators led to the alarm going off in any tested conditions after 120 s without triggered breaths. Auto-triggering was seldom found and easily worked out, except for in one ventilator.
Conclusions:
An appropriate alarm setting and combination of VT and TI would allow the majority of the tested ventilators to be used for mouthpiece ventilation without alarm activation.
More Related Videos
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
04:16Inspiratory Muscle Training as an Adjunct to the Treatment of Weaning Failure in Critically Ill Patients: A Practical Guide
Published on: January 30, 2026
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...