Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Vector Components in the Cartesian Coordinate System01:29

Vector Components in the Cartesian Coordinate System

30.6K
Vectors are usually described in terms of their components in a coordinate system. Even in everyday life, we naturally invoke the concept of orthogonal projections in a rectangular coordinate system. For example, if someone gives you directions for a particular location, you will be told to go a few km in a direction like east, west, north, or south, along with the angle in which you are supposed to move. In a rectangular (Cartesian) xy-coordinate system in a plane, a point in a plane is...
30.6K
Curvilinear Motion: Rectangular Components01:23

Curvilinear Motion: Rectangular Components

1.5K
Curvilinear motion characterizes the movement of a particle or object along a curved path, notably evident when envisioning a car navigating a winding road. If the car starts at point A, its position vector is established within a fixed frame of reference, where the ratio of the position vector to its magnitude signifies the unit vector pointing in the position vector's direction.
As the car advances, its position evolves over time. Quantifying the car's velocity involves computing the...
1.5K
Average and Instantaneous Velocity Vectors01:12

Average and Instantaneous Velocity Vectors

9.1K
To calculate other physical quantities in kinematics, the time variable must be introduced. The time variable not only allows us to state where an object is (its position) during its motion, but also how fast it’s moving. The speed at which an object is moving is given by the rate at which the position changes with time. For each position, a particular time is assigned. If the details of the motion at each instant are not important, the rate is usually expressed as the average velocity v.
9.1K
Vector Transformation in Rotating Coordinate Systems01:16

Vector Transformation in Rotating Coordinate Systems

2.9K
Consider a vector rotating about an axis with an angular velocity, such that its tip sweeps a circular path.
2.9K
Curvilinear Motion: Normal and Tangential Components01:27

Curvilinear Motion: Normal and Tangential Components

1.1K
When a car traverses a curved road, its motion can be elucidated by breaking it down into tangential and normal components. The car-centric coordinates attached to the vehicle move with it.
The positive direction of the t-axis aligns with the increasing position of the car along the curved path, denoted by the unit vector ut. Simultaneously, the n-axis, perpendicular to the t-axis, dissects the curved path into differential arc segments, each forming the arc of a circle with a radius of...
1.1K
Vector Algebra: Method of Components01:08

Vector Algebra: Method of Components

20.5K
It is cumbersome to find the magnitudes of vectors using the parallelogram rule or using the graphical method to perform mathematical operations like addition, subtraction, and multiplication. There are two ways to circumvent this algebraic complexity. One way is to draw the vectors to scale, as in navigation, and read approximate vector lengths and angles (directions) from the graphs. The other way is to use the method of components.
In many applications, the magnitudes and directions of...
20.5K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Quantification Beyond Binary of MR FLAIR Hyperintensity Lesions in Acute Ischemic Stroke of Unknown Time Since Onset.

Diagnostics (Basel, Switzerland)·2026
Same author

Ultrasound-Based Techniques for Visualization of Dermal Microvasculature: A Scoping Review.

Diagnostics (Basel, Switzerland)·2026
Same author

An ultrasound-guided biopsy technique for obtaining supraclavicular brown fat biopsies and preadipocytes.

Scientific reports·2025
Same author

Accuracy of detecting critical findings using abbreviated brain MRI scan protocols as a prerequisite for AI-driven on-the-fly scan protocol adaptation.

European journal of radiology·2025
Same author

Simulation in Obstetric and Gynecologic Ultrasound Training: Design and Implementation Considerations.

Journal of clinical medicine·2025
Same author

Human lymph node microvascular imaging using a fast contrast-free super-resolution ultrasound technique.

Scientific reports·2025
Same journal

Theoretical Foundations of the Echo Envelope Statistical Modeling: A Tutorial.

IEEE transactions on ultrasonics, ferroelectrics, and frequency control·2025
Same journal

Practical Demonstrations of FR3-Band Thin-Film Lithium Niobate Acoustic Filter Design.

IEEE transactions on ultrasonics, ferroelectrics, and frequency control·2025
Same journal

Real-Time Heterogeneous Helical Wave Spectrum Method for Transabdominal Passive Acoustic Mapping.

IEEE transactions on ultrasonics, ferroelectrics, and frequency control·2025
Same journal

Cascaded Plane Wave Ultrasound Velocity Vector Imaging: In Vivo Feasibility in Carotid Arteries.

IEEE transactions on ultrasonics, ferroelectrics, and frequency control·2025
Same journal

Quantitative Acoustic Attenuation Scanning Using a Phase-Insensitive Ultrasound Computed Tomography System.

IEEE transactions on ultrasonics, ferroelectrics, and frequency control·2025
Same journal

FPGA-Accelerated CNN Reconstruction for Low-Power Sparse-Array Ultrasound Imaging.

IEEE transactions on ultrasonics, ferroelectrics, and frequency control·2025
See all related articles

Related Experiment Video

Updated: Mar 28, 2026

Echo Particle Image Velocimetry
16:31

Echo Particle Image Velocimetry

Published on: December 27, 2012

15.2K

Convex array vector velocity imaging using transverse oscillation and its optimization.

Jørgen Arendt Jensen, Andreas Hjelm Brandt, Michael Bachmann Nielsen

    IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
    |December 17, 2015
    PubMed
    Summary
    This summary is machine-generated.

    This study optimized vector flow imaging using transverse oscillation (TO) on convex arrays. The enhanced method significantly reduced velocity estimation errors in simulations and real-world measurements, improving diagnostic accuracy.

    More Related Videos

    High-speed Particle Image Velocimetry Near Surfaces
    11:59

    High-speed Particle Image Velocimetry Near Surfaces

    Published on: June 24, 2013

    34.0K
    Meso-Scale Particle Image Velocimetry Studies of Neurovascular Flows In Vitro
    08:00

    Meso-Scale Particle Image Velocimetry Studies of Neurovascular Flows In Vitro

    Published on: December 3, 2018

    8.9K

    Related Experiment Videos

    Last Updated: Mar 28, 2026

    Echo Particle Image Velocimetry
    16:31

    Echo Particle Image Velocimetry

    Published on: December 27, 2012

    15.2K
    High-speed Particle Image Velocimetry Near Surfaces
    11:59

    High-speed Particle Image Velocimetry Near Surfaces

    Published on: June 24, 2013

    34.0K
    Meso-Scale Particle Image Velocimetry Studies of Neurovascular Flows In Vitro
    08:00

    Meso-Scale Particle Image Velocimetry Studies of Neurovascular Flows In Vitro

    Published on: December 3, 2018

    8.9K

    Area of Science:

    • Ultrasound medical imaging
    • Biomedical engineering
    • Fluid dynamics

    Background:

    • Accurate blood flow estimation is crucial for diagnosing vascular diseases.
    • Traditional ultrasound methods face limitations in quantifying complex flow patterns.
    • Vector Flow Imaging (VFI) offers advanced capabilities for detailed flow analysis.

    Purpose of the Study:

    • To develop and validate an optimized transverse oscillation (TO) method for vector flow imaging using convex array probes.
    • To improve the accuracy and reduce bias in estimating blood flow velocity.
    • To assess the performance of the optimized TO method through simulations and experimental measurements.

    Main Methods:

    • Utilized a 3-MHz, 192-element convex array probe for simulations and measurements.
    • Implemented an optimization routine to adjust the lateral oscillation period based on depth.
    • Evaluated performance using simulated parabolic velocity profiles at various beam-to-flow angles and depths.
    • Conducted in vivo scans on healthy volunteers' livers.

    Main Results:

    • Optimization reduced the energy ratio from -17.1 dB to -22.1 dB.
    • Simulated data showed a decrease in standard deviation from 8.81% to 7.4% and bias from -15.93% to 0.78% at 90°.
    • Experimental measurements yielded a bias of -0.93% and standard deviation of 6.37% at 87°.
    • In vivo liver scans demonstrated realistic velocity estimates and consistent flow patterns.

    Conclusions:

    • The optimized TO method significantly enhances the accuracy of vector flow imaging with convex arrays.
    • This technique provides reliable and precise blood flow velocity estimations in both simulated and in vivo settings.
    • The improved VFI method holds promise for more accurate clinical diagnosis of vascular conditions.