Comparison of cardiac output measurements using transpulmonary thermodilution and conventional thermodilution
Takaharu Itami1, Yusuke Endo2, Kiwamu Hanazono1
1Veterinary Teaching Hospital, Graduate School of Veterinary Medicine, Hokkaido University, Sapporo, Hokkaido, Japan.
Objective:
To evaluate the agreement between cardiac output (CO) values obtained using a transpulmonary thermodilution technique (TPTDCO) and conventional thermodilution technique (TDCO) in anaesthetized dogs with fluid overload.
Study Design:
Prospective experimental study.
Animals:
Six healthy Beagle dogs aged 7-8 years.
Methods:
Dogs were anaesthetized with sevoflurane in oxygen, and catheters were inserted for TPTDCO and TDCO measurement. After instrumentation, baseline CO was measured using each technique at a central venous pressure (CVP) of 3-7 mmHg. Dogs were subsequently administered lactated Ringer's solution and 6% hydroxyethyl starch to induce fluid overload. CO measurements were obtained using each technique at CVP values of 8-12 mmHg, 13-17 mmHg, 18-22 mmHg and 23-27 mmHg. Agreements between CO measurements obtained with the respective techniques were analysed using Dunnett's test, Pearson's correlation coefficient and Bland-Altman analysis.
Results:
Thirty pairs of CO values were obtained, ranging from 1.45 L minute(-1) to 4.69 L minute(-1) for TPTDCO and from 1.30 L minute(-1) to 4.61 L minute(-1) for TDCO. TPTDCO and TDCO values correlated strongly (r(2) = 0.915, p < 0.001). The bias and mean relative bias between TPTDCO and TDCO were 0.26 ± 0.30 L minute(-1) (limits of agreement - 0.29 to 0.81 L minute(-1) ) and 9.7%, respectively.
Conclusions And Clinical Relevance:
TPTDCO and TDCO measurements obtained in anaesthetized dogs during fluid overload exhibited good agreement. Accordingly, transpulmonary thermodilution provides an accurate measurement of CO in dogs with fluid overload.
More Related Videos
08:12Invasive Hemodynamic Monitoring of Aortic and Pulmonary Artery Hemodynamics in a Large Animal Model of ARDS
Published on: November 26, 2018
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Related Concept Videos
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...