Personal exposure to static and time-varying magnetic fields during MRI procedures in clinical practice in the UK
Evridiki Batistatou1, Anna Mölter2, Hans Kromhout3
1Centre for Occupational and Environmental Health, University of Manchester, Manchester, UK.
Background:
MRI has developed into one of the most important medical diagnostic imaging modalities, but it exposes staff to static magnetic fields (SMF) when present in the vicinity of the MR system, and to radiofrequency and switched gradient electromagnetic fields if they are present during image acquisition. We measured exposure to SMF and motion-induced time-varying magnetic fields (TVMF) in MRI staff in clinical practice in the UK to enable extensive assessment of personal exposure levels and variability, which enables comparison to other countries.
Methods:
8 MRI facilities across National Health Service sites in England, Wales and Scotland were included, and staff randomly selected during the days when measurements were performed were invited to wear a personal MRI-compatible dosimeter and keep a diary to record all procedures and tasks performed during the measured shift.
Results:
98 participants, primarily radiographers (71%) but also other healthcare staff, anaesthetists and other medical staff were included, resulting in 149 measurements. Average geometric mean peak SMF and TVMF exposures were 448 mT (range 20-2891) and 1083 mT/s (9-12 355 mT/s), and were highest for radiographers (GM=559 mT and GM=734 mT/s). Time-weighted exposures to SMF and TVMF (GM=16 mT (range 5-64) and GM=14 mT/s (range 9-105)) and exposed-time-weighted exposures to SMF and TVMF (GM=27 mT (range 11-89) and GM=17 mT/s (range 9-124)) were overall relative low-primarily because staff were not in the MRI suite for most of their shifts-and did not differ significantly between occupations.
Conclusions:
These results are comparable to the few data available from the UK but they differ from recent data collected in the Netherlands, indicating that UK staff are exposed for shorter periods but to higher levels. These data indicate that exposure to SMF and TVMF from MRI scanners cannot be extrapolated across countries.
Insights
MRI staff in the UK experience lower overall exposure to static magnetic fields (SMF) and time-varying magnetic fields (TVMF) due to shorter exposure times, though peak levels can be high. These findings differ from international comparisons, highlighting country-specific variations in MRI safety assessments.
Area of Science:
- Medical Imaging Physics
- Occupational Health and Safety
- Biomedical Engineering
Background:
- Magnetic Resonance Imaging (MRI) is a crucial diagnostic tool.
- MRI environments pose occupational exposure risks from static magnetic fields (SMF) and time-varying magnetic fields (TVMF).
- Understanding staff exposure is vital for safety and international comparisons.
Purpose of the Study:
- To quantify personal exposure levels to SMF and TVMF in UK MRI staff.
- To assess the variability of exposure across different roles and facilities.
- To enable comparison of UK exposure data with international findings.
Main Methods:
- Measurements were conducted across 8 MRI facilities in the UK.
- 98 participants, including radiographers and other healthcare professionals, were randomly selected.
- Personal MRI-compatible dosimeters and staff diaries were used to record exposure and activities.
Main Results:
- Average peak SMF and TVMF exposures were 448 mT and 1083 mT/s, respectively.
- Time-weighted exposures were relatively low (SMF: 16 mT, TVMF: 14 mT/s) due to limited time in the MRI suite.
- Radiographers showed higher peak exposures, but time-weighted exposures did not significantly differ by occupation.
Conclusions:
- UK MRI staff exposure levels are comparable to limited UK data but differ from international studies, particularly the Netherlands.
- UK staff experience shorter exposure durations but potentially higher peak field strengths.
- Exposure levels to SMF and TVMF from MRI scanners are not universally comparable across countries.
Related Concept Videos
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Induction
A...
Induced Electric Fields: Applications