Rapid Identification of Gram Negative Bacteria from Blood Culture Broth Using MALDI-TOF Mass Spectrometry
Published on: May 28, 2014
Optimized Use of the MALDI BioTyper System and the FilmArray BCID Panel for Direct Identification of Microbial
B Fiori1, T D'Inzeo1, A Giaquinto1
1Institute of Microbiology, Università Cattolica del Sacro Cuore, Rome, Italy.
Abstract:
Despite the current reliance on blood cultures (BCs), the diagnosis of bloodstream infections (BSIs) can be sped up using new technologies performed directly on positive BC bottles. Two methods (the MALDI BioTyper system and FilmArray blood culture identification [BCID] panel) are potentially applicable. In this study, we performed a large-scale clinical evaluation (1,585 microorganisms from 1,394 BSI episodes) on the combined use of the MALDI BioTyper and FilmArray BCID panel compared to a reference (culture-based) method. As a result, the causative organisms of 97.7% (1,362/1,394) of the BSIs were correctly identified by our MALDI BioTyper and FilmArray BCID-based algorithm. Specifically, 65 (5.3%) out of 1,223 monomicrobial BCs that provided incorrect or invalid identifications with the MALDI BioTyper were accurately detected by the FilmArray BCID panel; additionally, 153 (89.5%) out of 171 polymicrobial BCs achieved complete identification with the FilmArray BCID panel. Conversely, full use of the MALDI BioTyper would have resulted in the identification of only 1 causative organism in 97/171 (56.7%) of the polymicrobial cultures. By applying our diagnostic algorithm, the median time to identification was shortened (19.5 h versus 41.7 h with the reference method; P < 0.001), and the minimized use of the FilmArray BCID panel led to a significant cost savings. Twenty-six out of 31 microorganisms that could not be identified were species/genera not designed to be detected with the FilmArray BCID panel, indicating that subculture was not dispensable for a few of our BSI episodes. In summary, the fast and effective testing of BC bottles is realistically adoptable in the clinical microbiology laboratory workflow, although the usefulness of this testing for the management of BSIs remains to be established.
More Related Videos
11:25Preparation of a Blood Culture Pellet for Rapid Bacterial Identification and Antibiotic Susceptibility Testing
Published on: October 15, 2014
09:07Author Spotlight: Accelerating Diagnostic Accuracy with Direct Identification of Gram-Negatives from Blood Culture Bottles
Published on: May 24, 2024
Related Concept Videos
Automated Microbial Diagnostics
MALDI-TOF Mass Spectrometry
Methods of Classification and Identification
Modern Molecular Taxonomy