The Contribution of Pin End-Cup Interactions to Clot Strength Assessed with Thrombelastography
1From the Department of Anesthesiology, University of Arizona, Tucson, Arizona.
Abstract:
Viscoelastic methods have been developed to assess the contribution of plasma proteins and platelets to coagulation in vitro to guide clinical transfusion therapy. One of the cardinal precepts of determining clot strength is making sure that the viscoelastic technique includes complete exposure of the plastic pin in the testing chamber with the fluid analyzed so as to assure maximal interaction of the cup wall with the pin surface. However, the various contributions of the pin surface area to final clot strength have not been investigated. That is, it is not clear what is more important in the in vitro determination of clot strength, the surface area shared between the cup and pin filled with fluid or the final viscoelastic resistance of the gel matrix formed. Thus, the purpose of this investigation was to determine the clot strength when only the tip of the pin was engaged with plasma thrombus and to compare these values with clot strength values obtained when the pin was completely in plasma. After determining the minimal amount of plasma required to cover a pin tip in a thrombelastographic system (30 μL), clot strength (elastic modulus, G) was determined in plasma samples of 30 or 360 μL final volume (n = 12 per condition) after tissue factor activation. The G value with 30 μL volume was 1057 ± 601 dynes/cm (mean ± SD; 95% confidence interval, 675-1439 dynes/cm), which was (P = 0.0015) smaller than the G value associated with 360-μL sample volumes, that was 1712 ± 48 dynes/cm (confidence interval, 1681-1742 dynes/cm). In conclusion, these data demonstrate that clot strength is not determined by a simple ratio of surface area of pin and cup to volume of sample, but rather strength is importantly influenced by the viscoelastic resistance of the fluid assessed.
More Related Videos
09:38A Microfluidic Flow Chamber Model for Platelet Transfusion and Hemostasis Measures Platelet Deposition and Fibrin Formation in Real-time
Published on: February 14, 2017
09:19In Vitro Microfluidic Disease Model to Study Whole Blood-Endothelial Interactions and Blood Clot Dynamics in Real-Time
Published on: May 24, 2020
Related Concept Videos
Clot Retraction and Fibrinolysis
Formation of the Platelet Plug
As the injured blood vessel contracts, endothelial cells undergo contraction, revealing collagen fibers in the basement membrane and underlying connective tissue. Furthermore, the plasma membrane of endothelial cells becomes adhesive, preparing the site for platelet adhesion. Platelets...
Extrinsic and Intrinsic Pathways of Hemostasis
The Extrinsic Pathway
The extrinsic pathway of coagulation is typically initiated by tissue damage that exposes blood to tissue factor (TF), a protein released by the damaged tissue cells outside the blood vessels—this interaction with TF triggers biochemical reactions involving specific clotting factors. The key player here is Factor VII, which...
Coagulation
During the coagulation phase, clotting factors, or procoagulants, play a vital role in initiating and progressing the coagulation cascade. This cascade is a series of reactions...
Intracellular Signaling Affects Focal Adhesions
Some...
Anticoagulant Drugs: Low-Molecular-Weight Heparins