Performance demonstration of 4πβ(LS)-γ coincidence counting system for standardization of radionuclides with complex
D B Kulkarni1, R Anuradha1, Leena Joseph1
1Radiological Physics & Advisory Division, Bhabha Atomic Research Centre, Mumbai 400085, India.
Abstract:
A standardization of (134)Cs and (131)I was carried out in order to demonstrate the performance and applicability of the 4πβ(LS)-γ coincidence counting system for standardization of radionuclides with complex decay scheme. The coincidence analyzer, capable of analyzing coincidence between beta and two gamma windows simultaneously, was developed and used for the standardization. The use of this dual coincidence analyzer has reduced the total experimental time by half. The activity concentrations obtained using the 4πβ(LS)-γ coincidence counting system, a 4πβ(PC)-γ coincidence counting system, and the CIEMAT/NIST method are in excellent agreement with each other within uncertainty limits and hence demonstrates its performance for standardization of radionuclides decaying with complex decay scheme. Hence use of this 4πβ(LS)-γ coincidence counting system can be an alternative method suitable to standardize radionuclides with complex decay scheme with acceptable precision.
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay:
Nuclear Stability
To hold positively charged protons together...
Radioactive Decay and Radiometric Dating
Radioactivity and Nuclear Equations
A nuclide of an element has a specific number of protons and...
Isotopes and Radioisotopes
An isotope containing...