Quantitative analysis of pasireotide (SOM230), a cyclic peptide, in monkey plasma using liquid chromatography in
Yunlin Fu1, Wenkui Li1, Jimmy Flarakos1
1Department of Drug Metabolism and Pharmacokinetics, Novartis Institutes for BioMedical Research, One Health Plaza, East Hanover, NJ 07936, USA.
Abstract:
A novel liquid chromatographic method with tandem mass spectrometric detection (LC-MS/MS) for the determination of Pasireotide (SOM230) was developed and validated with a dynamic range of 0.5-250ng/ml using 50μl of monkey plasma. SOM230 and the internal standard, [M+6]SOM230, were extracted from monkey plasma via μElution SPE. The acidified sample matrix was loaded onto the preconditioned Waters SPE plate for further processing. The analyte was eluted from the SPE plate using freshly prepared elution solvent, which was followed by dilution and LC-MS/MS analysis. By eliminating a step of evaporation of the SPE eluent, instead, injecting the eluent after a simple dilution, compound loss due to non-specific binding to the 96-well materials was prevented. Furthermore, freshly prepared elution solution was found a key to optimal extraction recovery of the analyte from monkey plasma. The optimal chromatographic separation was achieved on an Atlantis dC18 (50×2.1mm, 5μm particle size) column using gradient elution with a total analysis cycle time approximately 4min per injection. The mobile phases were water containing 0.5% acetic acid and 0.05% trifluoroacetic acid (TFA) (mobile phase A) and acetonitrile containing 0.5% acetic acid and 0.05% TFA (mobile phase B). The incorporation of TFA (0.05%, v/v) and acetic acid (0.5%, v/v) into the mobile phases was accompanied by the improved chromatography and minimized carryover due to the HPLC column. The current method was validated for specificity, sensitivity, matrix effect, recovery, linearity, accuracy and precision, dilution integrity, batch size and stability. The accuracy and precision for the LLOQs (0.5ng/ml) were within ±5.6% bias and ≤7.8% CV, respectively. From the intra-day and inter-day evaluations, the precision of the other QC samples (1.5, 7.5, 75 and 190ng/ml) ranged from 2.7 to 4.9% CV and the accuracy (% bias) from -1.3 to 7.3%, respectively. Additional assessment of incurred sample reanalysis (ISR) was conducted to demonstrate the ruggedness and robustness of the assay method. The validated method was successfully implemented to support a toxicity study in monkeys administered with 5 and 30mg of SOM230 in a single intramuscular injection of a long acting release (LAR) formulation.
More Related Videos
08:08Quantitative Mass Spectrometric Profiling of Cancer-cell Proteomes Derived From Liquid and Solid Tumors
Published on: February 27, 2015
12:34Two-dimensional Gel Electrophoresis Coupled with Mass Spectrometry Methods for an Analysis of Human Pituitary Adenoma Tissue Proteome
Published on: April 2, 2018