The ARTUS device: the first feasibility study in human cadavers.

Tim A Ludwig1, Philip Reiss, Marco Wieland

  • 1Department of Urology, University Hospital Hamburg-Eppendorf, Germany.

The Canadian Journal of Urology
|December 22, 2015
PubMed
Summary

The ARTUS artificial urinary sphincter was successfully implanted in human cadavers, demonstrating feasibility. This new device offers a potentially safer alternative for preventing urethral damage during implantation.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Thulium fiber laser versus Holmium laser enucleation of the prostate: 24-month outcomes from a prospective randomized non-inferiority trial.

World journal of urology·2026
Same author

Rates and Predictors of Persistent LUTS Medication Use After Laser Enucleation of the Prostate.

Lower urinary tract symptoms·2026
Same author

Robust Localization and Tracking of VRUs with Radar and Ultra-Wideband Sensors for Traffic Safety.

Sensors (Basel, Switzerland)·2026
Same author

Persistent structural alteration of the extended limbic system in adolescents with childhood maltreatment.

European psychiatry : the journal of the Association of European Psychiatrists·2026
Same author

Lower urinary tract symptoms after laser enucleation of the prostate in patients with and without preoperative indwelling catheter.

International urology and nephrology·2025
Same author

Place of Death of Cancer Patients Treated at a German Comprehensive Cancer Center.

Palliative medicine reports·2025

Area of Science:

  • Urology
  • Medical Devices
  • Surgical Innovation

Background:

  • A novel electro-mechanical artificial urinary sphincter, ARTUS, has been developed.
  • ARTUS utilizes a unique sequential pressure mechanism to potentially mitigate urethral damage.

Purpose of the Study:

  • To assess the feasibility of implanting the ARTUS device in human cadavers.
  • To evaluate the surgical technique and basic operational parameters of the ARTUS device.

Main Methods:

  • ARTUS device implantation was performed in six human cadavers (3 male, 3 female).
  • Data collected included operation time, cuff size, wire length, and any complications.
  • Surgical technique was compared to the established AMS 800 device.

Main Results:

Related Experiment Videos

  • The ARTUS device implantation was found to be straightforward, utilizing a technique similar to the AMS 800.
  • Mean operation time was 20 minutes, with average cuff sizes of 4.5 cm (male) and 6 cm (female).
  • A larger subcutaneous pouch was required for ARTUS compared to AMS 800, and the study was preclinical.

Conclusions:

  • The ARTUS artificial urinary sphincter can be implanted with relative ease.
  • Surgical and technical approaches are comparable to the AMS 800, facilitating adaptation by experienced surgeons.
  • The study confirms the potential for ARTUS as a viable alternative in artificial urinary sphincter implantation.