Comparative characteristics of endodontic drills
Related Experiment Video
Updated: Mar 28, 2026

Dynamic Navigation in Endodontics: Guided Access Cavity Preparation by Means of a Miniaturized Navigation System
Published on: May 5, 2022
Sonia Loska1, Marcin Basiaga1, Magdalena Pochrząst1
1Silesian University of Technology, Faculty of Biomedical Engineering, Department of Biomaterials and Medical Devices Engineering, Zabrze, Poland.
Abstract:
The work concerns the analysis of influence of the wear process of endodontic instruments on the mechanical and physico-chemical properties of the materials from which they are made. A detailed study of the microstructure, mechanical properties and corrosion resistance in the environment simulating work of the tool was conducted. The research was done for the new Mtwo endodontic files and after six times of use. In addition, the observations with a scanning electron microscope in order to reveal possible damage caused by the impact of a corrosive environment were carried out. The results showed that use of the tool by six times revealed damage at the edges of the blades and may cause a lack of continuity of the cutting line resulting in the uneven distribution of the resistive force which acts on the tool during operation.