How flatbed scanners upset accurate film dosimetry

L J van Battum1, H Huizenga, R M Verdaasdonk

  • 1Physics and Medical Technology, VU University Medical Center, Support Radiotherapy, PO Box 7057, 1007 MB Amsterdam, The Netherlands.

Summary

Flatbed scanners introduce a lateral scan effect (LSE) when digitizing Gafchromic film, impacting accurate dose verification. Light polarization is identified as the primary cause, necessitating corrections for precise dosimetry.