How flatbed scanners upset accurate film dosimetry
L J van Battum1, H Huizenga, R M Verdaasdonk
1Physics and Medical Technology, VU University Medical Center, Support Radiotherapy, PO Box 7057, 1007 MB Amsterdam, The Netherlands.
Flatbed scanners introduce a lateral scan effect (LSE) when digitizing Gafchromic film, impacting accurate dose verification. Light polarization is identified as the primary cause, necessitating corrections for precise dosimetry.
Area of Science:
- Medical Physics
- Radiological Imaging
- Dosimetry
Background:
- Film dosimetry offers high spatial resolution for dose distribution verification.
- Flatbed scanners are cost-effective for digitizing irradiated film.
- Lateral Scan Effect (LSE) is a known disadvantage of flatbed scanners, altering readout along the scan axis.
Purpose of the Study:
- To investigate the causes of LSE in flatbed scanners digitizing Gafchromic film.
- To differentiate LSE origins from anisotropic light scattering, exploring crosstalk, optical path length, and polarization.
- To quantify the impact of LSE on dose measurements.
Main Methods:
- Investigated LSE on Epson 1680 Expression Pro and 10000XL scanners using Gafchromic EBT, EBT2, and EBT3 films.
- Examined crosstalk using neutral density filters and polarization effects using linear polarizers.
- Measured film optical density (OD) from 0.25 to 1.1 (0.2 to 9 Gy) in transmission mode using RGB channels.
Main Results:
- LSE magnitude depends on scanner and film type, increasing up to 14% at maximum lateral positions for 9 Gy.
- Crosstalk was significant (up to 2%) only in high-contrast regions for small fields.
- Optical path length effects caused up to 3% deviation in extreme lateral positions.
- Light polarization emerged as the main LSE contributor, varying across RGB channels.
Conclusions:
- Gafchromic EBT films digitized with flatbed scanners exhibit optical effects including LSE.
- Accurate dosimetry necessitates correction for LSE, requiring determination per color channel and delivered dose.
- Understanding and correcting for LSE is crucial for reliable film dosimetry verification.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Biological Effects of Radiation
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET