Reduced Silent Occlusions with a Novel Catheter Infusion Set (BD FlowSmart): Results from Two Open-Label Comparative
Michael Gibney1, Zhenyi Xue2, Monica Swinney3
11 Diabetes Care, BD Medical Affairs, Becton Dickinson and Company , Franklin Lakes, New Jersey.
Diabetes Technology & Therapeutics
|December 25, 2015
Summary
A new insulin infusion set significantly reduced silent occlusions, a common cause of unexplained hyperglycemia in insulin pump users. This novel design may improve insulin delivery accuracy and patient outcomes.
Area of Science:
- Biomedical Engineering
- Medical Device Technology
- Endocrinology
Background:
- Insulin pump therapy is associated with unexplained hyperglycemia.
- Silent occlusions, or insulin flow interruptions without alarms, are a potential cause.
- Silent occlusions may necessitate insulin infusion set replacement.
Purpose of the Study:
- To compare the incidence of flow interruption events between a novel dual-ported insulin infusion set (BD FlowSmart™) and an existing set (Quick-set®).
- To evaluate the effectiveness of the BD FlowSmart™ set in reducing silent occlusions and other flow interruptions.
Main Methods:
- Two studies were conducted using insulin diluent infusions in healthy adults.
- Study 1: Pilot study comparing BD FlowSmart™ and Quick-set® with manual or device-assisted insertion (n=25).
- Study 2: Larger study comparing BD FlowSmart™ and Quick-set® after manual insertions (n=60), focusing on flow interruption reduction.
Main Results:
- BD FlowSmart™ demonstrated significantly fewer silent occlusions compared to Quick-set® in both manual and mechanical insertions (Study 1).
- Overall flow interruption events were significantly reduced with BD FlowSmart™ in Study 2 (75% risk reduction).
- The percentage of time with flow interruption was consistently lower for BD FlowSmart™ sets across both studies.
Conclusions:
- The novel side-ported BD FlowSmart™ insulin infusion set significantly reduces flow interruptions, including silent occlusions.
- This reduction in flow interruptions suggests a potential improvement in insulin delivery accuracy.
- The findings indicate that the BD FlowSmart™ set may offer a more reliable alternative for insulin pump users.