Arthroscopically guided navigation for repair of acromioclavicular joint dislocations: a safe technique with reduced
Jan Theopold1, Bastian Marquass1, Nikolaus von Dercks1
1Department of Orthopaedics, Trauma and Plastic Surgery, University of Leipzig, Liebigstr. 20, 04103 Leipzig, Germany ; Devision of Arthroscopy, Joint Surgery and Sport Injuries, University of Leipzig, Liebigstr. 20, 04103 Leipzig, Germany.
Background:
Accuracy evaluation of navigated image free placement of double cortical fixation buttons for coracoclavicular tunnel position in comparison to conventional drill guide based placement.
Methods:
Twenty-six patients with acute acromioclavicular joint instability were included in this non-randomized cohort study. All patients were treated with a Double- TightRope technique. In 13 cases the conventional drill guide based placement was used (group 1). In 13 patients surgery was performed as a navigated procedure with a fluoro-free optoelectronic system (group 2). The number of coracoclavicular drillings per patient (First pass accuracy; FPA (%)) was documented, the subcoracoidal position of the fixation buttons has been evaluated and graded as "intended position achieved (IPA)" or "intended position not achieved (IPnA)".
Results:
In group 1 drilling had to be repeated in four patients (30.8 %) to achieve proper placement of the subcoracoidal fixation buttons. 100 % first pass accuracy was observed in group 2 (p = 0.03). In group 1, the intended position of the subcoracoid buttons was not achieved (IPnA) in six patients (46.2 %). In group 2 all intended positions were achieved (p = 0.005).
Conclusion:
Arthroscopic controlled fluoro-free navigated coracoclavicular drilling for the repair of acromioclavicular joint dislocation has higher first pass accuracy in comparison to conventional drill guide based placement. Therefore the navigation enables a precise position of the drill holes, may reduce the risk of an iatrogenic coracoid fracture and migration of fixation devices.
Trial Registration:
Local institutional review board No. 061-14-10032014.