Twiddler's syndrome in a patient with CRT-D device - A case report
Dheeraj D Bhatt1, Ripen K Gupta1, Upendra Kaul2
1Fortis Rajan Dhall Hospital, Sector B, Pocket 1, Vasant Kunj, New Delhi 110070, India.
Indian Heart Journal
|December 26, 2015
Abstract:
Twiddler's syndrome is a rare cause of pacemaker lead dislodgement. A 61-year-old female patient was implanted with a defibrillator capable of cardiac re-synchronization therapy (CRT-D); 10 months later, she presented with uneasiness and vibratory sensations in the chest. Fluoroscopy revealed rotation of the pulse generator along its longitudinal axis and dislodgement of all three leads. Diagnosis of Twiddler's syndrome was made.