A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
A NOVEL APPROACH FOR EFFECTIVE DOSE MEASUREMENTS IN DUAL-ENERGY CT
Brett J Mattison1,2, Giao B Nguyen2,3, Natalie Januzis1,2
1Medical Physics Graduate Program, Duke University, Durham, NC 27705, USA.
Abstract:
A novel method was presented for the effective dose (ED) measurement with metal-oxide-semiconductor field-effect transistor (MOSFET) detectors in dual-energy (DE) dual-source (DS) computed tomography (CT) scanner. This study demonstrated that the mean energy of the combined spectrum in dual-source computed tomography can be used to measure the ED. For validation, the MOSFET dose at the centre cavity of a CT dose index (CTDI) body phantom was compared with the dose measured with an energy-independent ion chamber (IC). A clinical abdomen/pelvis scan was performed with an adult anthropomorphic phantom, and ED was compared between the MOSFET method and the dose length product (DLP) method. The tissue doses in the CTDI phantom were 2.08±(2.70 %) with IC and 2.20±(4.82 %) cGy with MOSFET; the per cent difference was 5.91 %, and the t-test showed that there was no statistically significant difference. EDs for the abdomen/pelvis scans were 5.01±(2.34 %) mSv with MOSFET and 5.56 mSv with the DLP method.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Biological Effects of Radiation
X-ray Imaging
Imaging Studies III: Computed Tomography
Dose Size and Dosing Frequency: Determination Methods
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET