Experimental Study of Micro-Scale Taylor Vortices Within a Co-Axial Mixed-Flow Blood Pump
Fangjun Shu1, Ruijun Tian2, Stijn Vandenberghe1,3
1Department of Biomedical Engineering, Carnegie Mellon University, Pittsburgh, PA, USA.
Abstract:
Taylor vortices in a miniature mixed-flow rotodynamic blood pump were investigated using micro-scale particle image velocimetry (μ-PIV) and a tracer particle visualization technique. The pump featured a cylindrical rotor (14.9 mm diameter) within a cylindrical bore, having a radial clearance of 500 μm and operated at rotational speeds varying from 1000 to 12 000 rpm. Corresponding Taylor numbers were 700-101 800, respectively. The critical Taylor number was observed to be highly dependent on the ratio of axial to circumferential velocity, increasing from 1200 to 18 000 corresponding to Rossby numbers from 0 to 0.175. This demonstrated a dramatic stabilizing effect of the axial flow. The size of Taylor vortices was also found to be inversely related to Rossby number. It is concluded that Taylor vortices can enhance the mixing in the annular gap and decrease the dwell time of blood cells in the high-shear-rate region, which has the potential to decrease hemolysis and platelet activation within the blood pump.
More Related Videos
Related Concept Videos
Steady, Laminar Flow in Circular Tubes
Steady, Laminar Flow Between Parallel Plates
Bernoulli's Equation for Flow Along a Streamline
Couette Flow
Laminar and Turbulent Flow
Bernoulli's Equation for Flow Normal to a Streamline
The pressure difference depends on the fluid's velocity and radius of curvature. The pressure variation is minimal in flows with nearly straight streamlines. However, the...